27
Maxuum III Digital Electronic Indicator 2239028 Rev D
Press and release the Auto-Zero button for final zero setting.
The unit is now in standard measuring mode and ready to gage
parts.
NOTE: Some Mahr gages are furnished with separate instruc-
tions covering their use with Maxum III Indicators. Refer to
those instructions for specific steps for setup and operations.
NOTE: To Save battery life, the gage setup should permit the
Indicator to return to its Sleep Mode when not in use.
4.1.5 Setting Upper and Lower Tolerance Limits
Limits may be set to the resolution of the least digit anywhere
within each model's analog range. Both limits can be set on the
same side (unilateral) of zero or on opposite sides (bilateral) of
zero. Unit must be in Standard Operating Mode for this setting
procedure. Selection cannot be made while in TSP mode.
Any convenient method, such as a height stand or a bench
comparator, may be used to set limits. Once set, limits are
retained in memory, even when the Indicator is in its Sleep
mode, and remain set until changed. If no limits are needed,
move them outside the normally used portion of the measuring
range.
1.
Depress the spindle and press the Left Tolerance Limit
button or press the appropriate display
but-
ton to index the digital readout to the desired lower
limit (-) set point.
2.
Press and release the Left Tolerance Limit
button.
A blinking left arrow appears.
3.
Depress the spindle and press the Right Tolerance Limit
button or press the appropriate display
but-
ton to index the digital readout to the desired upper
limit (+) set point.
4.
Press and release the Right Tolerance Limit
button.
A blinking right arrow appears.
NOTE: To set limits, the measuring spindle must be positioned
within the analog range of the display. To use display buttons,
the Indicator must be in the measuring mode.
NOTE: To check limit settings, use display buttons to index the
fan and digital display through limit set points. Reset limits if
necessary. This can be used to set tolerance value more pre-
cisely. A single press and release of the display button moves
Содержание Maxum III
Страница 1: ...Instruction Manual Digital Indicator Document Part No 2239028 Rev D Maxum III 1 ...
Страница 40: ...40 Maxuum III Digital Electronic Indicator 2239028 Rev D Spindle Assembly Parts ...
Страница 55: ...55 Maxuum III Digital Electronic Indicator 2239028 Rev D 9 0 Quick Step Flow Chart ...