26
BA315001-en
Procedure
1 Switch the unit on and change to
"Measurement" mode.
2 With the unit in measurement mode,
switch off the unit at the master
switch so that the zero-balance
valve remains in the measurement
position.
3 Remove the sampling tubing from
the unit.