13
BA315001-en
5
Basic Settings
5.1
Display Language
Display: Menu > Set language
5.2
Time
MES: Otto or Diesel > Device diagnosis > Time/Date
Display: Menu > Set time
5.3
Date
The unit's date setting is configured before delivery. In accordance with official
calibration regulations, the date must not be altered by the end customer if the
unit is used in connection with the regular technical vehicle inspection as per
section 29 of the German Road Traffic Licensing Regulations (StVZO). Please
contact the customer service department as necessary.
MES: Otto or diesel > Device diagnosis > Date/Time
Display: Menu > Set date
6
Measurement Mode, Operating States and Settings
6.1
Initialisation and Warm-up Phase
When the unit is switched on, the message "Initialisation" is shown in the display
for roughly 10 seconds, while the PCB loads the firmware. As soon as the unit
shows the message "Warming up...", a four-gas measurement can be started.
This measurement is not affected by the warm-up phase.
The opacimeter takes roughly 2.5 minutes to warm up in preparation for
measurement. The warm-up phase is complete when the message "Warming
up..." disappears and the unit displays the main screen.
6.2
Leaktightness Test
A leaktightness test needs to be performed every day. If a measurement is started
before a leaktightness test has been performed, the unit automatically performs a
leaktightness test before the measurement is taken.
Observe the user prompts given in the display or the MES. To seal off the probe,
use the piece of tubing supplied. The test runs automatically, but can also be
triggered manually as required (for example, after the water extractor has been
cleaned).
MES: Otto or diesel > Device diagnosis > Calibration overview > Leaktightness
test