![Magtrol AN Series Скачать руководство пользователя страница 11](http://html2.mh-extra.com/html/magtrol/an-series/an-series_instruction-manual_3642441011.webp)
11
Magtrol AN 2000 C
Instruction Manual
11
2.2 – Programming instructions
Access to the programming mode
When power is applied to the instrument, the display briefly illuminates all
segments and LED's then shows the software version and finally enters in the
normal reading mode. Press
ENTER
to enter in the programming mode. The display
shows the indication "-Pro-" (fig. 11.1).
Exit from the programming mode without saving data
From any step of the program routines, a push of
ESC
returns the meter to the run mode. The instrument retains any change
made before exiting in this mode but does not save new data in the memory.
Save changes in the configuration
From the last step of each program menu, a push of
ENTER
returns the meter to the run mode keeping all changes in the
parameter list. The LED STORE illuminates while the new configuration is saved in the memory.
Guidelines on programming instructions
The programming software is divided into 6 modules. Each module is organized in several independently accessible menus and
each menu contains a list of parameters necessary to configure a specific function of the meter.
From the -Pro- stage, press repeatedly
to cycle around the existing modules: module 10 = Input configuration,
module 20 = display configuration, module 30 (if option installed) = setpoints, module 40 (if option installed) = analogical output,
module 50 (if option installed) = serial outputs and module 60 = logic functions. Press
ENTER
to accede selected module.
Fig. 11.1. PROG mode first step (-Pro- stage)
B
A
TARE
HOLD
LIMIT
MAX
MIN
DATA
DSP2
INP2
FLT
STORE
DSP1
INP1
2
1
4
3
TARE
RESET
LIMIT
MAX/MIN
ENTER
ESC
DATA
PROG
TEACH
PRG
RUN