34
Magtrol Model 3411 Torque Display
Chapter 6 – Mag.NET Commands
OPERA
TION
6.4
CALIBRATION AND MISCELLANEOUS COMMANDS
Command
Code
Function
Explanation
MODE#
User/Calibration mode
Values for #
0 User
1 Calibration/engineering
MODE 1
COMMANDS
INIT
Reset to factory defaults
Caution calibration will be lost. *POWER cycle
required after the command.
IOAOX.XXX
Write the value to DAC
Write the value to the Speed DAC. Volts are
assumed.
X.XXX floating point voltage value 3 decimal
10.000 to -10.000
CAL
Command to start A/D
Calibration
Response is ZERO
ZERO
Command to accept applied
zero
Response is FS=X.XXX prompting the user to
apply Full Scale
FS=X.XXX
Command to set the Positive
Full Scale (gain)
User enters X.XXX (meter reading)
Response is FN=-X.XXX
FN=X.XXX
Command to set the Negative
Full Scale (gain)
User enters X.XXX (meter reading)
Response is CAL COMPLETE
CALS
Command to start frequency
Calibration
Response is FREQ=XXXXX.XX prompting the
user to apply frequency
FREQ=
X.XXX
Command to set the frequency
correction
User enters X.XXX (meter reading) in Hz
Response is CAL COMPLETE
CALDAC
Command to start DAC
Calibration
Response is READ ZERO
ZDAC=X.XXX Command to set applied zero
User enters X.XXX (meter reading)
Response is ZERO OK?
ZERO OK
Command to accept DAC zero Response is FSDAC=X.XXX
FSDAC=
X.XXX
Command to set the Positive
Full Scale (gain)
User enters X.XXX (meter reading)
Response is FS OK?
FS OK
Command to accept DAC gain Response is CAL COMPDAC
OV
Reads Voltage
Reads voltage applied to Torque input mV.
Output sign follows the TM INVERT setting.
SPS#
Speed selection
Values for # are:
0 = Digital TachA output on pack panel BNC
1 = Analog speed output on pack panel BNC