31
Magtrol Model 3411 Torque Display
Chapter 6 – Mag.NET Commands
OPERA
TION
6.2
SETUP COMMANDS
Command
Code
Function
Explanation
AF#
Sets filter
Values for # are:
0 = None 4 = 20 Hz
1 = 2 Hz 5 = 50 Hz
2 = 5 Hz 6 = 100 Hz
3 = 10 Hz
M0
Front panel lockout
Front panel lockout, keys will not function.
M1
Front Panel enable
Front Panel enable, keys will function.
SPS#
Speed selection
Values for # are:
0 = Digital TachA output on pack panel BNC
1 = Analog speed output on pack panel BNC
SPSFxx.xx
Analog speed scale factor
Values for xx.xx range from 1 to 999,999.
TR
Resets Tare
Resets tare to calibrated 0 (zero).
TS
Sets Tare.
Reads current torque and uses as tare value.
UE#
Sets encoder units to #
Values for # range from
1 to 99999
UP#
Sets the Power Units.
Values for # are:
0 = hp
1 = W
2 = kW
UR#
Sets display torque units to #.
Values for # are:
0 = oz•in 6 = mN•m
1 = oz•ft 7 = cN•m
2 = lb•in 8 = N•m
3 = lb•ft 9 = kN•m
4 = g•cm 10 = daN•m
5 = kg•cm
Torque unit conversion defaults to Nm if out of
range.
UT#
Sets torque scale to #
Values for # range from 0.01 to 999999
INVERT#
Negates the incoming Voltage
(Torque signal)
Values for #
0 Non Inverted (default value)
1 Inverted
USER#
Selects configuration number
Values for #
1,2,3,4
NOTE:
If a USER SETUP is changed via
the USB or Ethernet it is volatile until SAVE
command is sent.
SAVE
Save set up
Save set up to current user configuration
SERIAL?
Returns Magtrol assigned
serial number
Содержание 3411
Страница 1: ...Model 3411 Torque Display User s Manual...