Magtrol WB/PB Dynamometers
Chapter 3 – Cooling
3.3
CLOSED-CIRCUIT COOLING SYSTEMS
The closed-circuit cooling system uses water-to-air or water-to-water heat exchangers, or a chiller.
This system is used when the water supply is restricted or when the water cannot be used for cooling
without prior treatment/filtering.
Note:
It is recommended to use water specific additives to avoid
proliferation of living organisms, and to protect against corrosion
and mineral deposits.
3.3.1
e
xAmPle
A closed-circuit cooling system simply requires a water tank in which the heat exchange will take
place. The ideal operating temperature of the dynamometer is maintained by a temperature-regulated
valve on the dynamometer cooling water outlet.
TORQUE
POWER
SPEED
Cold water
inlet
Pump with
water level
regulation
Water Tank
Valve regulated by the
dynamometer outlet water
temperature
Figure 3–2 Closed-circuit Cooling Example