45
Magtrol WB/PB Dynamometers
Chapter 2 – Installation / Mounting
2.1.4
WB/PB 115 S
erieS
WB/PB 115 Series Dynamometers are fitted with a mechanical stator assembly locking device
protecting the load cell during transport. After installing the dynamometer on the test bench, proceed
as follows and refer to
figure 2–4
to unlock the system:
1. Loosen the two nuts
.
2. Unscrew the two screws
with three turns.
3. Tighten the two nuts
.
Note: The screws and nuts for overload protection
and
are adjusted
in the factory (refer to
Chapter 6 – Calibration
for possible
adjustments).
Figure 2–4 WB/PB 115 Load Cell Locking System