Magstim 2nd Generation Coils
Operating Manual
© The Magstim Company Limited
27
MOP04-EN-03
Guidance and Manufacturer’s Declaration –
Electromagnetic Immunity
The Magstim Rapid
2
System is intended for use in the electromagnetic environment specified below.
The customer or the user of the equipment should assure that it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic environment
- guidance
Electrostatic
discharge (ESD)
IEC 61000-4-4
± 6 kV contact
± 8 kV air
Meets requirement
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Electrical fast
Transient/ burst
IEC 61000-4-11
± 2 kV for power
supply lines
± 1 kV for input/
output lines
Meets requirement
(Resetting of the MEP
Pod may be experienced
during single shot mode
under these conditions.)
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
± 1 kV differential
mode
± 2 kV common
mode
Differential mode
compliant.
Common mode compliant
to ± 1kV only.
Anti-surge protection needs to be
incorporated into the main supply
to the equipment if surge
protection is to be guaranteed.
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines.
IEC 61000-4-11
<5 %
U
T
(>95 % dip in
U
T
)
for 0,5 cycle
40 %
U
T
(60 % dip in
U
T
)
for 5 cycles
70 %
U
T
(30 % dip in
U
T
)
for 25 cycles
<5 %
U
T
(>95 % dip in
U
T
)
for 5 sec
Meets requirement
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the Magstim Rapid
2
System
requires continued operation
during power interruptions, it is
recommended that the Magstim
Rapid
2
System be powered from
an uninterruptible power supply.
Power
Frequency
(50/60Hz)
Magnetic Field
IEC 61000-4-8
3 A/m 50Hz
Meets requirement
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment
Radiated RF
Immunity
EN61000-4-3
80MHz
–
2.5GHz
2Hz 80%
amplitude
modulation
Coil adapter P/N 3110-00
and Cooled Coil P/N
3530-00 between 80MHz
and 100MHz may
produce system errors
setting system into safe
mode.
Equipment should only be used in
the vicinity of other equipment
compliant with EN60601-1-2.
Conducted RF
Immunity
EN61000-4-6
0.15MHz
–
80MHz
2Hz 80%
amplitude
modulation
Coil adapter P/N 3110-00
and Cooled Coil P/N
3530-00 between 30MHz
and 80MHz may produce
system errors setting
system into safe mode.
Equipment should only be used in
the vicinity of other equipment
compliant with EN60601-1-2.
NOTE
U
T
is the a.c. mains voltage prior to application of the test level