See Sections 2.3 & 2.4 of Installation and Operating
Manual 58-603 for more detailed application information
and limitations.
5.1.2 Environmental
See Section 3.7 of Installation and Operating Manual
58-603 for environmental limitations.
5.2
Skill Level of Personnel
Personnel following the procedures of this safety manual
should have technical expertise equal to or greater than that
of a qualified instrument technician.
5.3
Necessary Tools
Following are the necessary tools needed to carry out the
prescribed procedures:
• Open-wrenches or adjustable wrench to fit the process con-
nection size and type.
•
Antenna
2" (50 mm)
•
Transmitter
1
1
⁄
2
" (38 mm)
•
Torque wrench is highly desirable
• Flat-blade screwdriver
• Digital multimeter
• 24 VDC power supply, 23 mA minimum
5.4
Storage
The device should be stored in its original shipping box
and not be subjected to temperatures outside the storage
temperature (-50 to +80 °C) shown in the PULSAR Model
R86 Installation and Operating Manual and associated
specifications.
6
58-651 Pulsar
®
Model R86 SIL Functional Safety Manual