background image

Guided Wave Radar

Level Transmitter

Enhanced Model 705 with

F

oundation

fieldbus

digital output

705 software v3.x

F

oundation

fieldbus

operating Manual

Содержание Enhanced Model 705

Страница 1: ...Guided Wave Radar Level Transmitter Enhanced Model 705 with Foundation fieldbus digital output 705 software v3 x Foundation fieldbus operating Manual...

Страница 2: ...ection provided by equipment may be impaired Notice of Copyright and Limitations Copyright 2016 Magnetrol International All rights reserved Magnetrol Magnetrol logotype and Eclipse are regis tered tra...

Страница 3: ...l Mount 19 3 4 5 2 Remote Mount 19 4 0 Function Blocks 20 4 1 Overview 20 4 1 1 Universal Fieldbus Block Parameters 20 4 2 Resource Block 21 4 3 Transducer Block 24 4 3 1 Transducer Block Parameters 2...

Страница 4: ...e varying 4 20 mA current a digi tal communications scheme such as FOUNDATION fieldbus considers the two wires as a network The network can carry many process variables as well as other information Th...

Страница 5: ...sult A FOUNDATION fieldbus system allows for mul tiple variables to be brought back from each device to the control room for additional trending and reporting 3 Maintenance The self diagnostics residi...

Страница 6: ...tween the power supply in the safe area and the device in the hazardous area H1 also supports the Fieldbus Intrinsically Safe Concept FISCO model which allows more field devices in a network The FISCO...

Страница 7: ...g wiring and configuring the Eclipse level transmitter These procedures are intended for experi enced installers of electronic level measurement instruments See Complete Installation Section 3 0 for d...

Страница 8: ...nection or flange bolts NOTE Leave the plastic protective cap in place until ready to install the transmitter Do not use sealing compound or TFE tape on probe connection to transmitter as this connect...

Страница 9: ...nce with all regulations and codes 1 Remove the cover of the upper wiring compartment of the transmitter 2 Attach a conduit fitting and mount the conduit plug in the spare opening Pull the power suppl...

Страница 10: ...ext step of the configuration program the default password is 1 7 After entering the last value allow 10 seconds before removing power from the transmitter The following configuration entries are the...

Страница 11: ...future reference when ordering parts Model Number Serial Number 3 2 Electrostatic Discharge ESD Handling Procedure MAGNETROL electronic instruments are manufactured to the highest quality standards Th...

Страница 12: ...t the process connection size and type Coaxial probe 11 2 38 mm twin rod probe 17 8 47 mm transmitter 11 2 38 mm A torque wrench is highly desirable Flat blade screwdriver Fieldbus compatible power su...

Страница 13: ...Pressure probe Model 7xP High Pressure probe Model 7xR Overfill probe Model 7xS Steam probe and Model 7xT Interface probe require added clearance Process temperature pressure dielectric and viscosity...

Страница 14: ...rs etc a bare tank wall parallel to the probe is acceptable 3 4 2 1 To install a rigid twin rod probe Make sure the process connection is at least 2 NPT or a flanged mounting Make sure that there is a...

Страница 15: ...evices 2 b Loosen the two 10 32 set screws 3 on both securing devices using a 3 32 2 5 mm hex wrench and slide the securing devices off of the probe c Slide the TFE weight off of the probe d Cut and r...

Страница 16: ...ed connections tighten the hex nut of the probe process connection For flanged connections tighten flange bolts Probe can be stabilized by placing into a non metallic cup or bracket at the bottom of t...

Страница 17: ...ds Probes The Model 7x2 and 7x5 Bulk Solids probes are designed for a 3000 lb 1360 kg pull down force for use in applications such as sand plastic pellets and grains It is offered with a maximum 75 fo...

Страница 18: ...nches of coating from the two cables f Slide the weight back on to the probe g Reinstall the two cable clamps and tighten h Enter the new probe length inches or cm in software 3 4 4 4 To install a Mod...

Страница 19: ...ction for wiring configuration and viewing When the transmitter is facing the desired direction use a 11 2 38 mm wrench to tighten the universal connection on the transmitter to 15 ft lbs A torque wre...

Страница 20: ...Block Parameters The following are general descriptions of the parameters common to all blocks Additional information for a given parameter is described later in that specific block section ST_REV st...

Страница 21: ...Block The RESOURCE block contains data specific to the Enhanced Model 705 transmitter along with some information about the firmware NOTE The Resource Block has no control function MODE_BLK Must be i...

Страница 22: ...interval between copies of non volatile NV parameters to NV memory NV mem ory is only updated if there has been a significant change in the dynamic value and the last value saved will be available fo...

Страница 23: ...he static data in the block BLOCK_ALM Block Alarm is used for configuration hardware connection or system problems in the block The cause of any specific alert is entered in the subcode field The firs...

Страница 24: ...ollowed by these additional required parameters UPDATE_EVT Update Event an alert generated by a write to the static data in the TRANSDUCER block Another important parameter found later in the TRANS DU...

Страница 25: ...on the nameplates attached to the transmitter and probe PROBE_LENGTH is shown in SENSOR_UNITs LEVEL_OFFSET Enter the distance from the probe tip to the desired 0 reference in PROBE_UNITs The accept ab...

Страница 26: ...f the probe 10 inches above the bottom of the tank The user wants the 0 point at 24 inches and the 100 point at 60 inches as referenced from the bottom of the tank When the Eclipse transmitter is moun...

Страница 27: ...arameters are WINDOW CONVERSION_FACTOR and SCALE_OFFSET WINDOW is used to adjust for the variations in the analog section of the Eclipse TDR measurement engine CON VERSION_FACTOR and SCALE_OFFSET are...

Страница 28: ...y supplied when simulate is enabled When simulate is disabled the simulate value and status track the actual value and status XD_SCALE The high and low scale values engineering units code and number o...

Страница 29: ...The block alarm is used for all configuration hardware connection failure or system problems in the block ALARM_SUM The current alert status unacknowledged states unreported states and disabled states...

Страница 30: ...t to Direct with improper OUT_SCALE The AI block uses the STATUS_OPTS setting and the TRANSDUCER PV LIMIT value to modify the AI PV and OUT QUALITY Damping Filter is a feature of the AI block The PV_F...

Страница 31: ...e will be displayed subject to limitations nec essary for a 6 character display 999999 Value 99999 Representative examples are shown below Because the Model 705 transmitter has four Analog Input block...

Страница 32: ...algorithm as Backward or Bi linear BAL_TIME The specified time for the internal working value of bias to return to the operator set bias BKCAL_IN The analog input value and status for another blocks B...

Страница 33: ...he alarm limit used to detect the HI alarm condition HI_PRI The priority of the HI alarm IN The connection for the PV input from another block LO_ALM The LO alarm data LO_LIM The setting for the alarm...

Страница 34: ...allowed SP_LO_LIM The lowest SP value allowed SP_RATE_DN Ramp rate for downward SP changes SP_RATE_UP Ramp rate for upward SP changes STATUS_OPTS Allows one to select options for status handling and...

Страница 35: ...ne Transmitter Display All MeasType selections 4 PrbModel select User Select the type of probe used Select from 7xa x 7xB x 7xd x 7xE x 7xF E 7xF F 7xF x 7xF 4 7xF P 7xG x 7xJ x 7xK x 7xL x 7xM x 7xn...

Страница 36: ...ersion 22 DispFact select None Select Yes to display factory parameter menus 23 History Status None Diagnostic Display to view present status and recent exceptions 24 Run Time xxxx x h None Diagnostic...

Страница 37: ...7xK Distance in mm from fiducial to user reference point 44 Snsr Val xxx x su None Distance to the target relative to the sensor reference point 45 CalPtLo xxx x su Superuser The lower calibrated poi...

Страница 38: ...to OOS the TRANSDUCER BLOCK will still output level but the QUALITY will be shown as Bad and the SUB_STATUS is Out of Service When the Enhanced Model 705 measurement cycle fails to find a valid output...

Страница 39: ...the Analog Input block The Simulate feature is typically used to exercise the operation of an AI block by simulating a TRANSDUCER block input This feature can not be activated without the placement of...

Страница 40: ...near very top of probe Check for buildup or obstructions near top of probe Select higher dielectric range Check for condensation in probe connection Add Blocking Distance LEVEL and AI OUTPUT values P...

Страница 41: ...t Value Mode OK 15 0x00008000 Good Non Specific Not Limited Mode User Access 7 0x00000080 Bad OOS Not Limited Mode Factory Access 5 0x00000020 Bad OOS Not Limited Fault Default Params 28 0x10000000 Ba...

Страница 42: ...robe Model is proper for the dielectric of the medium 4 Consult factory EOP Probe Length End of Probe signal from a dry probe is out of range 1 Ensure the Probe Length is entered correctly 2 Set Diele...

Страница 43: ...cable shield can be capacitively grounded in multiple places to improve EMC protection Ensure all devices are on the live list and the schedule has been downloaded Ensure that the Resource Block then...

Страница 44: ...up G T4 Class III Type 4X 705 52XX XXX Non Incendive Class I Div 2 Groups A B C D 705 52XX XXX Suitable for Class II Div 2 Group G T4 Class III Type 4X atEX 705 52XX AXX Intrinsically Safe II 1G EEx i...

Страница 45: ...7 2 1 agency Specifications Foundation fieldbus System 57 640 Eclipse Guided Wave Radar Transmitter FOUNDATION fieldbus 45...

Страница 46: ...56T6 0 20 copper optional 316 stainless steel Cable Entry 3 4 NPT and M20 46 57 640 Eclipse Guided Wave Radar Transmitter FOUNDATION fieldbus Environment Operating Temperature 40 to 175 F 40 to 80 C D...

Страница 47: ...nce Model 705 Specifications will degrade with Model 7xB 7xD and 7xP probes and or fixed threshold configuration Top 24 inches of Model 7xB probe 1 2 inches 30 mm Specification for top 48 inches of si...

Страница 48: ...plosion Proof Z030 9151 003 FISCO Z030 9151 004 O ring Viton 012 2201 237 Consult Factory for alternative O ring materials Housing cover without glass 004 9193 003 Housing cover with glass GP IS 036 4...

Страница 49: ...e Intrinsically Safe ATEX JIS EEx ia IIC T4 C Integral Explosion Proof ATEX EEx d ia IIC T4 must be ordered with Conduit Connection Codes 0 and 1 D Remote Explosion Proof ATEX EEx d ia IIB T4 must be...

Страница 50: ...316L stainless steel NACE construction K 316 316L stainless steel probe and process connection ASME B31 1 specifications model 7xS only N Enlarged coaxial probe 316 316L stainless steel probe 2 minimu...

Страница 51: ...500 ANSI Raised Face Flange 23 1 150 ANSI Raised Face Flange 24 1 300 ANSI Raised Face Flange 25 1 600 ANSI Raised Face Flange 27 1 900 1500 ANSI Raised Face Flange 28 1 2500 ANSI Raised Face Flange 3...

Страница 52: ...orque Tube Flange UT 31 2 600 Masoneilan Proprietary Carbon Steel Torque Tube Flange UU 31 2 600 Masoneilan Proprietary 316 Stainless Steel Torque Tube Flange UV 31 2 600 Masoneilan Proprietary Carbon...

Страница 53: ...5 StrapTbl Pt05 56 STRAPPING_TABLE_POINT06 StrapTbl Pt06 57 STRAPPING_TABLE_POINT07 StrapTbl Pt07 itEM PaRaMEtER naME PaRaMEtER LaBEL 58 STRAPPING_TABLE_POINT08 StrapTbl Pt08 59 STRAPPING_TABLE_POINT0...

Страница 54: ...l Process Medium Tag Electronics Serial TROUBLESHOOTING Probe Serial Working Value Non Working Value Level Volume optional Interface optional Interface Volume opt Probe Model Probe Mount Measurement T...

Страница 55: ...eet Copy blank page and store calibration data for future reference and troubleshooting Item Value Value Value TROUBLESHOOTING Working Value Non Working Value FidTicks FidSprd Fid Type HF Cable Fid Ga...

Страница 56: ...will be determined on the basis of the applicability of our warranty No claims for misapplication labor direct or consequen tial damage will be allowed Return Material Procedure So that we may effici...

Отзывы: