3/5
LH71/LH72 Segmented error compensation Setting manual
Create compensation data for the changing points narrowed down to 32 points or less from the error curve measured
in
step 1.
* Since the reference point position is 0, it is not included in the compensation points.
Compensation point (P)
P1
P2
Reference
point position
P3
P4
P5
Count value of counter unit(
l
)
-500.0000
-125.000
0.000
200.000
414.000
700.000
Lase display value
-499.9860
-125.008
0.0000
200.032
414.008
700.020
Compensation amount(ε)
0.014
-0.008
0.0000
0.032
0.008
0.020
Error curve compensation data for the above 5 changing points
(Unit: mm)
Changing point:
- Point where the slope of the error curve changes
- Point where the inclination is reversed
Number of points that can be input:
up to 32 points
Compensation amount:
up to
±
600μm
Number the changing point in order from the smallest (or minus) side of the count value of the measurement unit.
Distance traveled
based on the
reference point
(
l
)
When the changing point is 5 points
Compensation point input number
Compensation amount at compensation point
Position data of compensation point
Creation of compensation data
STEP 2
Reference point position: 0
Number the changing point in order from the smallest (or minus) side of the count value of the measurement unit. If the start position of
the segmented error compensation is position 1 and the end position is position N, the position N should be 32 or less.
The maximum compensation value for the entire changing point is within
±
600μm. If it exceeds the
±
600μm range, the segmented
error compensation function cannot be used.
NOTE:
- The segmented error compensation function is applied to the range from the start position (ε 1) to the end position (ε n).
- Input compensation data cannot be handled below the unit of input resolution of the measurement unit.
Error amount (
ε
)
l 1
l 2 l 3
l 4
l 5
ε
1
ε
2
ε
3
ε
4
ε
5
P 1
P 2
P 3
P 4
P 5