LH71/LH72 Segmented error compensation Setting manual

To use the segmented error compensation function, it is necessary to measure the amount of error with the distance actually traveled
with respect to the count value of the measuring unit. To measure the actual distance traveled, use a laser interferometer or the like. The 
segmented error function can be used in combination with the scale with built in reference point. The base point for the segmented error 
compensation is based on the reference point position of the scale (0 mm).

Changing points 

Distance

Error amount

P1

l 1

ε

1

P2

l 2

ε

2

P3

l 3

ε

3

P4

l 4

ε

4

P5

l 5

ε

5

Error amount 

ε

ε

M

ε

C

ε

M

Master (laser interferometer, etc.) 

ε

C

: Count

value of display unit

Actually move the stage (or object) and collect the entire error data. Create an error curve from the data and narrow 
down the changing points (compensation points) to 32 points or less. If there are two or more axes, consider using 
the average value for each axis or narrowing down to the data of frequently used positions.

The measurement range that can correct the segmented 
error compensation is the range of 7 digits displayed.

3. Setting of segmented error compensation function

The range of segmented error compensation depends on the minimum resolution 
used. In addition, the maximum number of points that can be corrected for section 
error is 32 points excluding the reference point. The maximum total compensation 
amount is 

±

600um. The count value of the measuring unit with respect to the actual 

distance traveled is affected by the direction of movement of the table and the 
change in attitude of the machine due to the movement of the stage.

Measurement unit with built in reference point

Reference point

Resolution

Correctable range

Compensation 

points

0.1um

±

999.9999 mm (App. 

±

1m)

Up to 32 points
(Excluding 
reference points)

0.5um

±

999.9995 mm (App.

±

1m)

1.0um

±

9999.999 mm (App.

±

10m)

10um

±

99999.99 mm (App. 

±

100m)

Drive range matrix on the X-Y axis plane

Reference point position: 0

A finer error curve can be obtained by making the 

sampling interval for acquiring error data finer.

Narrow down the changing points of the error curve to 32 

points or less. Number from 1 in order from the minus side 

or the smallest value. 

(Excluding the reference point position)

Measured error curve

Error curve of linear error

Laser interferometer

Moving

fixed

Laser-head

X axis

Stage

Y axis

If the slope of the error curve is a straight curve, 
the linear error compensation function may be 
sufficient instead of the segmented error 
compensation function.

When the changing point is 5 points

Y1

Y2

Y3

X1

X2

X3

+

+

Measurement of mechanical error

Sampling pitch

In the case of X and Y axes, 
6 measurements are required, 
even if at least the left, right, 
and center of each axis are 
taken into consideration.

STEP 1

Error amount

ε

Error amount

ε

Moving distance

Distance traveled based 
on the reference point

Distance traveled 
based on the 
reference point

(

l

)

Reference point position: 0

Error amount (

ε

)

l 1

l 2 l 3

l 4

l 5

ε

1

ε

2

ε

3

ε

4

ε

5

P 1

P 2

P 3

P 4

P 5

X

Y1

Data1

Y2

Data2

Y3

data3

X1

X2

X3

Y

data4 data5 data6