16 

 

MAGNAMED     

Ventilator Analyzer - 

VentMeter

  P/N: 1901509 

– Rev 05 

When  adjusted  in  1,  it  is  possible  to  see  a  complete  respiration  cycle  with  60  min

-1

 

(respiration rate) rate. Adjusting in 12, it is possible to see a complete respiration cycle with 5 
min

-1

 respiration rates. 

 

Flow Measurement Screens: 

 

 

 

By pressing the ENTER 

 key, in the 

Main Screen

, when the arrow is pointing to Flow, the display will 

show the 

Flow Measurement Screen

The F1 and F2 key allows navigation between flow screens, notice that as the F2 key is pressed the number 
at the superior right corner is incremented, indicating greater distance from the 

Main Screen

 

Flow Screen (1): 

Flow Numerical Values 

o

 

Central Screen 

 Instantaneous Flow 

o

 

Inferior Screen 

 Minimum Flow and Maximum Flow 

 

Flow Screen (2): 

Flow Graph 

o

 

Central Screen 

 Flow Graph 

o

 

Inferior Screen 

 Instantaneous Flow Value 

 

Flow Screen (3): 

Graph’s Scale 

o

 

Max Flow 

 Maximum value of flow in the graph (adjusted from 10 to 150 L.min

-1

 with steps 

of 10 L.min

-1

o

 

Time 

  One  screen  sweep  time  (adjusted  from  1  to  15  seconds  with  steps  of  1  second). 

When  adjusted  in  1,  it  is  possible  to  see  a  complete  respiration  cycle  with  60  min

-1

 

(respiration rate) rate. Adjusting in 12, it is possible to see a complete respiration cycle with 5 
min

-1

 respiration rates. 

 

Volume Measurement Screens: 

 

 

 
 

By pressing the ENTER 

 key, in the 

Main Screen

, when the arrow is pointing to Volume, the display 

will show the 

Volume Measurement Screen

The  F1  and  F2  key  allows  navigation  between  volume  screens,  notice  that  as  the  F2  key  is  pressed  the 
number at the superior right corner is incremented, indicating greater distance from the 

Main Screen

 

Volume Screen (1): 

Volume Numerical Values 

o

 

Central Screen 

 Instantaneous Volume 

o

 

Inferior Screen 

 Vmin: Minute Volume; Vexp: Expired Volume; Vinsp: Inspired Volume 

 

Volume Screen (2): 

Volume Graph 

o

 

Central Screen 

 Volume Graph 

o

 

Inferior Screen 

 Instantaneous Volume Value 

 

Volume Screen (3): 

Graph’s Scale 

Содержание VentMeter

Страница 1: ...INSTRUCTION MANUAL Ventilator Analyzer VentMeter Part Number 1901509 Rev 05...

Страница 2: ...cnologia M dica S A Rua Santa M nica 801 831 Bairro Capuava CEP 06715 865 Cotia SP Brazil Tel 55 11 4615 8500 E mail magnamed magnamed com br Website www magnamed com br MAGNAMED Tecnologia M dica S A...

Страница 3: ...possibility of injury death or other serious adverse reaction due to the misuse of the equipment Caution Warns the user about the possibility of equipment failure associated with the use or misuse Lik...

Страница 4: ...uate lung ventilators performance It should not be used to monitor patient ventilation in Intensive Care Units This manual should be completely read to safely and correct use of the equipment and to g...

Страница 5: ...nently charged o Recharge the batteries completely after use or after a long period of storage Notes Dispose of removed parts in accordance with the institution s parts and parts disposal protocol and...

Страница 6: ...ECTRICAL POWER CONNECTION 25 7 3 OXYGEN SENSOR OPTIONAL 25 7 4 BEFORE USE 26 7 4 1 Initial Procedures 26 8 VENTILATOR TEST SAMPLES 27 9 TROUBLESHOOTING 31 10 PARTS AND ACCESSORIES 33 11 PREVENTIVE MAI...

Страница 7: ...interpret ventilation curves Basic Characteristics AC Input 100 240 VAC 50 60 Hz Internal battery with intelligent charger LCD Display to show pressure curve flow curve volume curve and monitored valu...

Страница 8: ...ssure Plateau Pressure Minimum Pressure Static Compliance Dynamic Compliance Airway Resistance Beside these monitored data by the flow sensor it is possible to measure the oxygen concentration of the...

Страница 9: ...ntMeter P N 1901509 Rev 05 4 Part Identification 4 1 Parts and Components The following items are supplied integrating the equipment Quantity Description Picture 01 VentMeter Ventilator Analyzer 01 Ki...

Страница 10: ...lator Analyzer VentMeter P N 1901509 Rev 05 Quantity Description Picture 01 Testing bag set 01 Handbag 01 CD Instruction Manual and PulmoTrend Software for PC basic version 01 Source AC DC 90 2source...

Страница 11: ...cated by a highlighted text in the screen Cal and Next 3 LCD Display Show graphs and monitored parameters 4 UP Key or Increment Key This key increments or UP ARROW to allow navigation within the scree...

Страница 12: ...r Caution Connect the tubing according to the picture shown in the front panel of the VentMeter 9 Proximal Flow Sensor Connector Proximal flow sensor pressure measurement port connector 10 PC Serial R...

Страница 13: ...Padroniza o The internal battery of the Ventilator Analyzer should be always charged and ready to use in eventual electrical power failure to achieve it this cable should be kept connected to the main...

Страница 14: ...14 MAGNAMED Ventilator Analyzer VentMeter P N 1901509 Rev 05 5 Command and Adjustment...

Страница 15: ...he letter at superior left corner indicates which flow sensor is selected A Adult Sensor I Infant Sensor N Neonatal Sensor Pressure Measurement Screens The F1 and F2 keys allows navigation between pre...

Страница 16: ...ph s Scale o Max Flow Maximum value of flow in the graph adjusted from 10 to 150 L min 1 with steps of 10 L min 1 o Time One screen sweep time adjusted from 1 to 15 seconds with steps of 1 second When...

Страница 17: ...uence with up to 3 parameters selected Screen sequence with up to 6 parameters selected By pressing the ENTER key in the Main Screen when the arrow is pointing to Monitor the display will show the Mon...

Страница 18: ...f the arrow and the ENTER key selects or inhibits the parameter to be shown in the monitor screen The selected parameters will be blinking and the inhibit ones will not To navigate between these param...

Страница 19: ...parameters Vexp Freq and FiO2 and Monitor Screen 2 with Vinsp in the middle of the screen Notice that in this situation the Monitor Screen 2 will show in its superior right corner the number 2 indica...

Страница 20: ...ree flow sensors The blinking flow sensor is the one selected To select another flow sensor use the UP and DOWN keys to reach the new sensor and press ENTER key to select it This new sensor will start...

Страница 21: ...nd 100 the arrow indicates which oxygen concentration will be used to calibrate the sensor If the arrow is pointing to 21 and the ENTER key is pressed then it will blink and the message Calibrating wi...

Страница 22: ...on of the ventilator to the patient Tinsp Inspiratory Time Is the inspiratory time value Positive Flow Texp Expiratory Time Is the expiratory time value Negative Flow I E I E Ratio Is the result of Ti...

Страница 23: ...Connect the flow sensor tubing according to the following figure 2 Connect the oxygen sensor to VentMeter using appropriate cable audio cable 3 Turn on the VentMeter and wait for the initialization sc...

Страница 24: ...01509 Rev 05 5 Prepare the respiratory circuits to test the ventilator Warning The respiratory circuit should be STERILIZED otherwise there is risk of contamination 6 Carry out the following connectio...

Страница 25: ...of the VentMeter only with the power of the internal battery it must be fully recharged and ready for further use If the VentMeter remains unplugged for more than one month the battery must be fully...

Страница 26: ...RP16P02 Tee Adaptor o MAXTEC R110P10 Flow Diverting 7 4 Before Use This inspection routine is intended to orient the user to carry out a simple procedure and quick equipment test before each use or a...

Страница 27: ...figure bellow b Calibrate the Ventilator Analyzer VentMeter c Calibrate or carry out the initialization procedures of the ventilator according to the manufacturer s recommendation d Adjust the lung s...

Страница 28: ...ximum Pressure Pplat Plateau Pressure PEEP Positive Expiratory End Pressure Evaluating adult ventilator in PCV Pressure Controlled Ventilation mode a Connect the system as shown in the figure bellow b...

Страница 29: ...25 cmH2O Frequency Respiration Rate 12 min 1 Inspiratory Time 1 7 s PEEP 5 cmH2O f See the curves Pressure x Time Flow x Time Volume x Time sample in the figure bellow g Monitor the following paramet...

Страница 30: ...30 MAGNAMED Ventilator Analyzer VentMeter P N 1901509 Rev 05 Sample 3 Test x Result Table Reference IEC 60601 2 12 Table 101...

Страница 31: ...nsducer 1 Find it and reconnect it firmly 2 Verify the simulation condition 3 Ask for technical service assistance No flow curve 1 Respiratory circuit disconnection 2 Simulated respiratory mechanics c...

Страница 32: ...ocedure according to chapter 6 3 Ask for technical service assistance FiO2 cell doesn t read correctly 1 Electrical cable connection to the oxygen cell 2 Pneumatic connection of the oxygen sensor 3 Ce...

Страница 33: ...LMOTREND BASIC 01 PC 7 2701886 POWER SUPPLY AC DC 12V 1A 90 TO 260 VAC 01 PC 8 3102183 CONNECTOR 90 15M 15F 01 PC 9 3201471 NEONATAL FLOW SENSOR POLICARBONATE 01 PC 10 3201472 PEDIATRIC FLOW SENSOR PO...

Страница 34: ...rical power cords LCD display readability Key operation F1 F2 UP ENTER DOWN Integrity of the flow sensor connection tubing Integrity of the flow sensors Integrity of the oxygen sensor cable optional A...

Страница 35: ...o 12VDC On Off Switch O2 Cell sensor OPTIONAL O2 Cell connection cable OPTIONAL One 20L min 1 flow sensor One 50L min 1 flow sensor One 150L min 1 flow sensor 12 1 2 Electrical Characteristics Item Pa...

Страница 36: ...mL or 3 of reading Infant Sensor INF 2 mL 10 to 400 2 5 mL or 5 of reading Neonate Sensor NEO 2 mL 1 to 100 0 1 2 mL or 5 of reading 7 Minute Volume Adult Sensor ADU L 0 1 to 50 0 0 001 0 18 L or 3 o...

Страница 37: ...he tolerance is calculated by frequencies 12 20 and 30 Rpm respectively to adult infant and neonate sensor The tolerance is a function of volume uncertainty multiplied by the frequency 4 When two tole...

Страница 38: ...vice installation operation or modifications carried out by non authorized or disqualified personnel The lack of sealing label or its rupture by non authorized or disqualified personnel voids this war...

Страница 39: ...be operated by qualified personnel Magnamed Tecnologia M dica S A Rua Santa M nica 801 831 Bairro Capuava CEP 06715 865 Cotia SP Brasil Tel Fax 55 11 4615 8500 E mail magnamed magnamed com br Website...