3
English
L-10 Coil
Operating Manual
PRODUCT SPECIFICATIONS
Waveform
AC
Coil Size
10.25 in / 26 cm ID
15 in / 38 cm OD
2.5 in / 6.3 cm wide
Field Strength
2,856 Ampere-Turns
Duty Cycle
25%, maximum on time 15 sec
Power Cord
12 ft / 4 m
Control Cord
5 ft / 1.5 m
Weight
115 V / 13.1 lb / 6.0 kg
230 V / 11 lb / 5.0 kg
Electrical
115 V / 60 Hz / 12 A
230 V / 50 Hz / 6 A
Environmental
32-120°F (0-49°C), up to 100% RH, up to 16,300 ft /
5,000 m altitude, non-condensing
Regulatory
Designed for mains +/- 10%, over voltage category
II, wet locations, pollution degree 2 environment
OPERATION
1.
Position the coil on a bench or clamp vertically in vise.
2.
Magnetic Particle Inspection
a.
Place part inside coil. Magnetic field direction is along the axis of coil (longitudinal).
b.
Press foot switch to energize. Apply magnetic particles to part. While energized,
blow off excess powder to reveal indications.
c.
Evaluate indications for size, shape, and relevant locations.
3.
Demagnetization
a.
Place part inside coil.
b.
Press foot switch to energize. Withdraw part from coil, up to 18 in / 45 cm distance,
while coil is energized.
ACCESSORIES
Part Number
Description
169799 (NA)
008M003 (EMEAR)
Pie Field Gage
008M004
Magnetic Flux Indicators, set of 5
2480
Field Indicator Gauge, 10 Gauss