MagCore Super Скачать руководство пользователя страница 13

12

MagCore® 

Super / 

MagCore® HF16

 Plus 

━ If the optic test result is 

over detection limit

Select to enable or disable the optic measurement 

function. (MagCore Super Only)

(Not available for 105, 201, 202, and 211)

Please confirm the parameters and press the Start 

button to begin the program.

Automated extraction process begins.

The extraction process is completed.

If you select the optic measurement funtion, you 

can browse, print or output the test result to a USB 

Flash Drive (please refer to the optic data chapter).

If the test result of the optic test shows the 

concentration is over detection ( DNA >300ng/μl / 

RNA > 240 ng/μl), the system will ask you to dilute.

A.Please record and save the 

current measured data.

Please select YES.

Please check the value for each sample.  If 

there is a value showing DNA >300 ng/

μl or RNA > 240 ng/μl, please record 

the number of the sample.  It means the 

sample needs to be diluted.  After recording 

the number of the slot, please insert the 

USB Flash Drive and press the Save button 

to save to current measured data.  After 

finishing the above steps, please press Next 

Step.

 

10

7

8

9

102

Program Code :

 Select Optic Measurement Function

Enable

Disable

Back

Press Start to Process

Back

201

Program Code :

Start

N/A

200

N/A

300

Enable

DNase : 

Sample Volume : 

Elution Buffer : 

Elution Volume : 

Optical Measurement : 

Program Completed

OK

Completed

Print Data

Review Data

OK

Program Completed

OK

Program Completed
10X Dilution ?

Print Data

Yes

No

Program Completed

OK

Program Completed
10X Dilution ?

Print Data

Yes

No

Note the samples which need dilution

Prev

Barcode

1234567890

> 300

2.05

2.364

1.154

ng / μL ( DNA )

260 / 280

A260

A280

2014 / 01 / 30  09:19:59

Date

Next

Next Step

Save

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16

Stop

Program Processing

201

Program Code :

Start

Remaining Time

Process Message

2014/01/01 23:59:59

10 / 75

D2 Lamp OFF

1

The values are DNA >300 ng/μl / 

RNA > 240 ng/μl

Please save the data in the USB Flash Drive 

and transfer to computer, or the data in 

the USB Flash Drive may be overwritten 

later.

File name: OpticsSampleData.cvs

(MagCore Super Only)

Содержание Super

Страница 1: ...UserInterface 10 EmergencyStop 10 StartPrograms 11 Iftheoptictestresultis 12 overdetectionlimit 12 Stop 14 Enablethebarcodefunction 14 Barcodescanningandediting 15 Optictestdata 15 Easymaintenance 16...

Страница 2: ...of the type of power supply at your lab consultthesellerorlocalpowercompany Forproductsintendedtooperatefrom batteryorothersources refertotheoperatinginstructions 11 Grounding or Polarization This pro...

Страница 3: ...edbytheClot Detectionfunctionoftheinstrument Clumpingofbeadsmayleadtolowyieldandpurityofnucleicacids cross contamination andinhibitionofdownstreamassays e g PCR Aftertherunhasfinished carefullyinspect...

Страница 4: ...oesnotwarrantthatsamplestreatedwithLysisBufferarecompletelyinactivatedand noninfectious Aftersampleprocessingiscompleted removeandautoclavealldisposableplastics Donoteat drinkorsmokeinthelaboratorywor...

Страница 5: ...ts Warning electricity Instructionmanualmustbereadbeforeoperating Consultinstructionsforuse Alternatecurrentinput Use byDate Nouseperiod Temperaturestorageconditions15 C 30 C Cataloguenumber productco...

Страница 6: ...0xH770 mm W27 95xD29 92xH30 31 inches NetWeight 76 5kg 168 68LB MagCoreSuper 68 5kg 151LB MagCoreHF16Plus Specification Operating Parameters Operating Environment Applications Whole Blood DNA 200 l 40...

Страница 7: ...vetteSet 10pcs X1 11 WasteBoxX1 12 BarcodeScannerX1 MagCoreSuperOnly Please refer to the machine size shown in the above figure to prepare an appropriate position for the machine and reservespaceforth...

Страница 8: ...alde16slots n mero1 16daesquerdaparaa direita 12345678910111213141516 Printer MagCoreSuperOnly TouchPanel FrontDoor TopServiceLid CoolingFans BackServiceLid OperationArea USBPort ScannerPort BarcodeSc...

Страница 9: ...etoputitintothewellifthe directionisnotcorrect 2 Putthe200 lSPtipintotheW4oftheT Rack Pleaseinsertthefrontendofthecartridgeintothespace belowthefixingplateoftheCartridgeRack 1 Putthetipintothecorrespo...

Страница 10: ...3 Testheater PleaserefertoHeaterTestchapterfortheinformationabout theoperation Step4 Testopticmodule MagCore SuperOnly PleaserefertoOpticTestchapterfortheinformationabout theoperation Whentheopticmodu...

Страница 11: ...e The above figureshowsthe function of the operation icons Start Maintenance Barcode Optics Data Please read the security information to correctly stop the machine whenanaccidenttakesplace When an eme...

Страница 12: ...ystemwillaskyouwhetherornot todeletepreviousbarcoderecord Ifthereisno barcoderecord youwillnotbeasked MagCore Superonly HF16Plusoptional Start Maintenance Barcode Optics Data 401 ProgramCode SelectElu...

Страница 13: ...rtthe USBFlashDriveandpresstheSavebutton tosavetocurrentmeasureddata After finishingtheabovesteps pleasepressNext Step 10 7 8 9 102 ProgramCode SelectOpticMeasurementFunction Enable Disable Back Press...

Страница 14: ...disappears Ifthefilesavedlasttimeislost you canretrievethisfileaccordingtothe instructionsofoptictestdatachapter OK Next Step Back 10XDilution 1 Take away and store the products which are NO need to...

Страница 15: ...esstheAbortbuttonifyouwanttoend thetest Duringtheabortionprocess moduleswillreturnto theiroriginalpositions Pleasedonotoperatethe machineduringtheprocess Themachinehasbeenstoppedandreset please presst...

Страница 16: ...code Optics Data Barcode Cancel Scan Print Data Delete All OpticalData Display Optical Data Print Optical Data Back Display Previous Optical Data Start Optical Measurement Print Previous Optical Data...

Страница 17: ...stonTest Easymaintenance 2 1 Start Maintenance Barcode Optics Data Stop O Ring Next Step Syringe O Rings 1 2 3 4 Maintenance Reject Tip Piston Test UV Heater Test Replace O Ring Motor Test Next Back M...

Страница 18: ...Well3ofT Rack 2 Pleaseadd1mlwatertotheSampleTubeandputitin theWell1oftheT Rack 3 PleaseselecttheTestW3Tip 4 Pleasecheckwhethertheliquidlevelsinalltipsare consistentwitheachother Iftheliquidlevelsareno...

Страница 19: ...ips are still on the pistons you have to press Reject Tip to remove thetipsfromthepistons PleaseselecttheRejectTipbutton Clean Piercing Needles 2 1 1 Stop PiercingNeedles Next Step Prev Maintenance Up...

Страница 20: ...ptheD2lamp Thescreenwillshowthemeasuredvaluesafter thetestisfinished YoucaninserttheUSBFlash DriveandthenpresstheSavebuttontostorethe currentdataaftertheUSBiconshowsinthestatus bar PleasetakeouttheUSB...

Страница 21: ...edles System Status Optic Test Alarm History Back Clear Save to USB AlarmMessage Back Date Time ErrorMessages Code AlarmMessage Back 0000001 ErrorCode Pleaserefertousermanualor contactyourlocalenginee...

Страница 22: ...methatyouchangedin Current Time afterthechangeisfinished System Status Prev Maintenance Update Program Vender Piercing Needles System Status Optic Test Alarm History Back SystemStatus Back System Time...

Страница 23: ...sfinished pleasepressOKtocheck whethertheUIversioniscorrect Pleasedumpthewastewhentheprogramis completed andwashthewasteboxeverytime Warning Pleaseconfirmthenameof UIversion Iftheversionisnot correct...

Страница 24: ...following figure and thenclosethecoveroftheprinter If the position of the thermal paper roll is not correct the red yellow lamp will blink Please press the Feed button to confirm the installation is...

Страница 25: ...pdated the following sections General Considerations Document Revision History Update to comply with the requirements of the IVD Regulation EU 2017 746 June 2023 EN MagCore Super HF16 Plus Operation M...

Страница 26: ...15F No 15 QiaoheRd ZhongheDist NewTaipeiCity235029 Taiwan TEL 886289121200FAX 86289121300...

Отзывы: