GA-21Plus
and
-
move the cursor to the sensor to be calibrated
-
select the gas type
Calibration screen will appear:
All currently installed electrochemical sensors are visible in this screen
The individual rows have the following meanings:
Cell
– Sensor type
Signal
– Current value of sensor's signal
Stored
– Value of stored signal
-
in the
GAS [ppm]
line, starts editing of the standard
gas concentration
to
-
enter the concentration of the standard gas [ppm] in
the line
GAS
41