![Madur GA-21 Plus Скачать руководство пользователя страница 32](http://html.mh-extra.com/html/madur/ga-21-plus/ga-21-plus_operating-manual_688625032.webp)
www.madur.com
Manual
GA-21Plus
5.4. Pressure/Draft measurements
It is possible to measure differential pressure with the
GA-21
plus
. To do so the
stubs PRESSURE (-) and (+) on the front plate are used. The measurement is performed
as follows:
-
select the main MENU
and
-
move the cursor to
PRESSURE
-
select the option. The following screen appears
The current pressure and gas temperature are displayed on the screen. The last
line shows the command
ZEROING
. By pressing this key the sensor is set to zero, e.g.
the current value will be taken as zero value..
Please be aware that the sensor has to be zeroed before the measurement.
-
ZEROING the sensor
Connect end of the sample hose to the
PR)
stub. The screen will dis-
play the measured pressure and the measured gas temperature.
-
leaves the option. The displayed value will be stored
until a new one is stored or either a printout of the
measured values is performed or they are stored as
a report.
32