5.6.2.5. Sensor calibration
Except for O2 every electrochemical sensor needs to be calibrated with calibra-
tion gas. To calibrate the sensors they are fed one by one with calibration gas of a known
concentration. After a few minutes the sensor supplies a stabilised signal which can be
stored as calibration value. The order of calibration does not effect the calibration.
Warning:
Any calibration screen value change is followed by loss of the original cal-
ibration settings. Thus the CALIBRATION option should be selected only when it is
to be carried out.
The standard gas concentration should be similar to that usually met during
measurement and it should not be less than 10% of the required measuring range of a
given sensor
The time the sensor is fed with standard gas should not be less than 4 min. After
each calibration the sensors should be flooded with ambient air for at least 15 minutes.
Calibration procedure:
and
-
move the cursor to
CALIBRATION W. GAS
-
select the option
-
carry on with the calibration
-
abort the calibration, move back to
SENSORS