4.3.5. Power supply
The
GA-21
plus
can be used with mains power supply or be operated from the
internal battery. A soot test can only be carried out when the mains supply is connected,
otherwise the heating of the probe holder will not function. The mains cable is attached to
the socket on the left of the keyboard. Switching the mains supply on or off when the in -
strument is in operation is not recommended. The analyser switches automatically to the
internal battery when the mains supply is disconnected. The LED on the front panel
shows which power supply is in use.
Red
-analyser is switched on and supplied from the battery. Mains supply
not present.
Green
-analyser switched off and mains supply present. The battery is char
ging.
Orange
-analyser switched on and mains supply present. The battery is char
ging.
The maintenance-free 12 V battery has a capacity of 2 Ah. A fully charged battery
will power the instrument for about five hours.
When the battery voltage drops below 11 V, three acoustic warnings and blinking
of the LED on the front panel will appear. The message
ERROR
appears at the top of the
screen and the
CONTROL LIST
contains the line
BATT LOW
.
The instrument switches off automatically if the battery voltage drops below 10.5
V.
The battery is charged by connecting the instrument to the mains power supply.
The state of charge can be seen during the initial calibration or read from the
screen
DEVICE INFO: