www.madur.com
Manual
GA12
Drawing 6.
Initial calibration screen.
The icon in the top-left corner of the screen shows the load of the battery. When
the initial calibration is done, the Parameter's screen is displayed. It allows to check / set
measurement parameters.
3.3. Screens
3.3.1. Results screens
To enter the Results screen (available from any other screen) press ESC or Res-
ults function keys. The drawing below shows an example Results screen:
Drawing 7.
Results screen
In the top part of the display there is an informative bar. The meaning of the sym -
bols, starting from the left:
1
- The first out four possible Results screen. To switch between
those screens use up / down arrow keys.
M61
- Informs about the the available memory for reports. M61 means
there is a space for another 61 reports.
F2
- Selected fuel. The analyser has 10 fuels stored in its memory – 6
are defined by producer, user can modify the remaining 4 fuels and their
parameters. Fuels are numbered fro 0 to 9.
9