www.madur.com
Manual
GA12
Drawing 12.
Measurement parameters screen
Parameters that user can adjust here:
Averaging time
– The analyser stores, in its internal buffer, the results every
2 seconds. However the results displayed on the screen are averaged for the
selected time. To select averaging time press Change function key or left /
right arrow keys. Available averaging times: 2, 10, 20, 30, 60, 120, 180
seconds.
Reference O2
– this is the oxygen value used for calculations of e.g. mass
concentrations of gases referred to oxygen. Use left / right arrow keys to
change the O2 reference value. Range of change 0..13%, step 1%. If, in the
PC software's setting, one of the following is selected:
Önorm fuels
or
Use
O2ref from fuel data
, then it is not be possible to change the O2 reference
value from the analyser's settings – the O2 reference value stored in fuel
parameters is used.
Fuel
– Currently set fuel – its parameters are used for all the calculations.
Fuel parameters are stored in the analyser from the PC software. From the
analyser's menu, user can select one of the stored fuels – use Change func-
tion key or left / right arrow keys. The fuel number is display on the right, the
fuel's name is displayed in the line beneath.
XL
measurement – sets the time of single XL measurement. Use Change
function key or left / right arrow keys to adjust it. Available XL measurement
times: 2s, 6s, 10s, 20s, 30s, 1min, 2min, 3min.
Fuel
– Currently selected fuel, its parameters are used for calculating values.
Parameters of fuels F0÷F5 are defined by producer and stored permanently
in the device and they cannot be changed. Fuels F6÷F9 are editable by user
from the PC software. To change the Fuel press Change function key or use
14