MADSEN Zodiac isintended for use in the electromagnetic environment specified below. The user of MADSEN Zodiac should ensure that it isused in such an
environment.
Emissionstest
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
MADSEN Zodiac uses RF energy only for its internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
MADSEN Zodiac issuitable for use in all environments, including domestic environments and
those directly connected to the public low-voltage power supply network that supplies buildings
used for domestic purposes.
Harmonic emissions IEC
61000-3-2
Not applicable
Voltagefluctuations/flicker
emissionsIEC 61000-3-3
Not applicable
Guidanceand manufacturer's declaration - electromagnetic immunity for all equipment and systems
MADSEN Zodiac isintended for use in the electromagnetic environment specified below. The user of MADSEN Zodiac should ensure that it isused in such an
environment.
Immunitytest
IEC 60601
test level
Compliancelevel
Electromagnetic environment - guidance
Electrostatic discharge(ESD)
IEC 61000-4-2
+/- 6 kV contact
+/- 8 kV air
+/- 6 kV contact
+/- 8 kV air
Floors should bewood, concrete or ceramic tile. If floors
are covered with synthetic material, the relative humid-
ity should be at least 30 %.
Electrical fast transient/burst
IEC 61000-4-4
+/- 2 kV for power supply lines
+/- 1 kV for input/output lines
+/- 2 kV for power supply lines
+/- 1 kV for input/output lines
Mains power quality should be that of a typical com-
mercialor hospital environment.
Surge IEC 61000-4-5
+/- 1 kV line(s) to line(s)
+/- 2 kV line(s) to earth
+/- 1 kV line(s) to line(s)
+/- 2 kV line(s) to earth
Mains power quality should be that of a typical com-
mercialor hospital environment.
Voltagedips, short inter-
ruptionsand voltage vari-
ations on power supply input
lines IEC 61000-4-11
<5% U
T
(>95 % dip in U
T
) for 0.5
cycle
40% UT (60 % dip in U
T
) for 5
cycles
70% U
T
(30 % dip in U
T
) for 25
cycles
<5% U
T
(>95 % dip in U
T
) for 5s
<5% U
T
(>95 % dip in U
T
) for 0.5
cycle
40% UT (60 % dip in U
T
) for 5
cycles
70% U
T
(30 % dip in U
T
) for 25
cycles
<5% U
T
(>95 % dip in U
T
) for 5s
Mains power quality should be that of a typical com-
mercialor hospital environment. If the user of the
MADSEN Zodiac requirescontinued operation during
power mains interruptions, it isrecommended that the
MADSEN Zodiac bepowered from an uninterruptible
power supply or a battery.
Power frequency
(50/60 Hz) magnetic field
IEC 61000-4-8
3A/m
3A/m
Power frequency magnetic fields should be at levels char-
acteristic of atypical location in atypical commercial or
hospital environment.
U
T
is the AC mains voltage prior to application of the test level.
Guidanceand manufacturer's declaration - electromagnetic immunity - for equipment and systems that are NOT life-supporting
Otometrics - MADSEN Zodiac
User Guide