Storing and handling
Temperature:
-20°C to +60°C (-4°F to +140°F)
Relative humidity:
< 90 %, non-condensing
Air pressure:
500 hPa to 1060 hPa
Dimensions (HxWxD)
Stand-alone version:
190 mm x 248 mm x 261 mm (7.5” x 9.8” x 10.3”)
PC-based version:
100 mm x 240 mm x 240 mm (3.9” x 9.4” x 9.4”)
Probe dimensions (HxWxD)
Quick Check probe:
28 mm x 22 mm x 100 mm (1.1” x 0.9” x 3.9”)
Diagnostic probe:
10 mm x 10 mm x 25 mm (0.4” x 0.4” x 1.0”)
Weight
Stand-alone version:
2.65 kg/5.85 lb
PC-based version:
1.65 kg/3.64 lb
Optional features (Stand-alone)
Printer:
Built-in printer. Prints 832 dot line/s on 112 mm paper width
2 cc coupler
Calibration
Equipment should be calibrated regularly according to EN 60645-5 and ANSI S3.39
Essential performance
MADSEN Zodiac has no essential performance and accordingly, the applicable requirements are as stated in the following:
1. Impedance/admittance as defined by EN 61027 Type 1, ANSI S3.39 Type 1.
2. Basic safety as defined by IEC 60601-1.
All information required by IEC 60601-1-2:2007, #5.2.2.1-#5.2.2.10 is available in the MADSEN Zodiac User Guide.
Standards
Safety:
IEC 60601-1, UL 2601-1, CAN/CSA - C22.2 NO 601.1-90
ANSI/AAMI ES60601-1 + AMD 1, CAN/CSA-C22.2 No. 60601-1
MADSEN Zodiac: EN 60601-1, Class II, externally powered, Type BF, IPX0
EMC:
EN 60601-1-2
Impedance/Admittance:
Clinical/Diagnostic: EN 60645-5 Type 1, ANSI S3.39 Type 1
Power supply:
Class I externally powered supply
Otometrics - MADSEN Zodiac
User Guide
46