Probe status and leakage
The light indicators
Light indicators light up in color in the probe to indicate different states. The control pad on the shoulder strap shows the
selected ear color at all times, if you are using the diagnostic probe.
Probe color
Status
Red
•
The right test ear has been selected
•
The device is in idle mode
Blue
•
The left test ear has been selected
•
The device is in idle mode
Green
•
The test is running
Yellow
•
Leak
OTOsuite indications
Color
Status
Green
•
During measurements, OTOsuite shows a green background to the online
values.
Probe leakage
If there is a probe leak during testing, this will be shown on the MADSEN Zodiac and OTOsuite screens.
If you are using the Quick Check probe, the leakage indication will remain until you remove the probe
from the ear and try again.
Probe blocked
If the probe is blocked during testing, this will be shown on the MADSEN Zodiac and OTOsuite screens.
If you are using the Quick Check probe, the blocked probe indication will remain until you remove the
probe from the ear and try again.
9.5
Sequence testing
A test sequence is a predefined set of automatic tests which can be performed automatically.
You can perform the immittance tests in a sequence:
•
Diagnostic sequence
Tympan Reflex Threshold (auto search) (+ Reflex Decay, if selected from the menu).
•
Screening sequence
Tympan Reflex Screening.