Editing results
Select a curve to edit the properties.
•
Delete
•
Swap Data Between Left and Right Ear
•
Set Threshold
•
Set No Response
•
Remove Threshold Information
12.7
Reflex Decay testing
This test is supported by MADSEN Zodiac Diagnostic and MADSEN Zodiac Clinical.
You can record a reflex decay measurement either as a separate measurement or as part of a diagnostic sequence. In
Sequence testing, the test automatically suggests the stimulus level, which is the reflex threshold level +10 dB.
In the following you will find the description for performing reflex decay testing as a single test. This means that the
sequence function is not enabled.
Warning
•
The Reflex Decay test is a supra-threshold test, where the stimulus levels are very high for an exten-
ded period of time. Make sure that there are no contraindications for performing the test.
Note
•
It is recommended that you perform a tympanometric test before making any acoustic reflex meas-
urement, and determine the acoustic reflex threshold before making a reflex decay measurement.
Note
•
If Reflex Decay is performed as part of a sequence, the following applies: The decay test stimulus level
pre-supposes that there is a reflex threshold available for the ear, stimulus type and stimulus ear (ipsi- or con-
tralateral) that is chosen for reflex decay testing. The decay test stimulus level will then automatically be set to
the threshold level +10 dB. If the reflex has not been determined, the decay measurement will be skipped.
98
Otometrics - MADSEN Zodiac
12 Testing with Zodiac Diagnostic or Clinical - PC-based
Содержание zodiac
Страница 1: ...MADSEN Zodiac Reference Manual Doc No 7 50 1610 EN 02 Part No 7 50 16100 EN...
Страница 21: ...4 2 Bottom view A Probe connections B Label Otometrics MADSEN Zodiac 21 4 Views of Zodiac...
Страница 156: ...jB jB App 2 5 2 G App 2 5 3 B G B G Y B G B G 156 Otometrics MADSEN Zodiac App 2 MADSEN Zodiac...
Страница 157: ...App 2 5 4 226 Hz 226 Hz 200 daPa Otometrics MADSEN Zodiac 157 App 2 MADSEN Zodiac...
Страница 206: ...152 151 151 152 150 152 206 Otometrics MADSEN Zodiac Index...