Starting the test
1. Use the dial to adjust the desired start pressure.
2. Press the
Start/Stop
button to initiate the test.
3. Use the dial to control the pump and create a manual tympanogram.
The derived measurement data in the tympanometric results table are calculated from the baseline reference data at
200 daPa. The results table accordingly remains empty, and the curve remains non-baseline compensated until you
have reached 200 daPa.
4. Press
Stop
to finish the measurement.
5. When you have finished testing one ear, switch ears, if needed.
Viewing and managing test results
6. To view tympanometry results, see
The Manual Tympanometry screen
.
7. When you have finished testing, see
Managing test results in MADSEN Zodiac
9.8.1
The Manual Tympanometry screen
A.
On-line values
B.
Tympanometry graph
C.
Tympanometry results
D.
On-screen selections
A. On-line values
The on-line values show the current settings and status.
•
Probe tone in Hz
•
Current equivalent volume/admittance (678, 800, 1000 Hz probe tone)
B. Tympanometry graph
The graph area shows the tympanometric curves and can rescale automatically to fit the curves.
•
Tympanometric curves
•
Pressure and admittance scales
•
Ear canal volume bar. The ear canal volume is shown to the right of the graph.
•
Norm area
Otometrics - MADSEN Zodiac
67
9 Testing with Zodiac Diagnostic or Clinical - Stand-Alone
Содержание zodiac
Страница 1: ...MADSEN Zodiac Reference Manual Doc No 7 50 1610 EN 02 Part No 7 50 16100 EN...
Страница 21: ...4 2 Bottom view A Probe connections B Label Otometrics MADSEN Zodiac 21 4 Views of Zodiac...
Страница 156: ...jB jB App 2 5 2 G App 2 5 3 B G B G Y B G B G 156 Otometrics MADSEN Zodiac App 2 MADSEN Zodiac...
Страница 157: ...App 2 5 4 226 Hz 226 Hz 200 daPa Otometrics MADSEN Zodiac 157 App 2 MADSEN Zodiac...
Страница 206: ...152 151 151 152 150 152 206 Otometrics MADSEN Zodiac Index...