Live baseline graph
The live baseline graph is an online admittance indicator which automatically monitors the admittance of the ear when
the probe is inserted.
Before you start a measurement, make sure that this baseline is as stable as possible. Fluctuations in the baseline will
have a destructive influence on the reflex measurements
To achieve a stable baseline, there are a number of factors influencing the stability of the baseline:
Patient behavior
•
The patient is completely still so that noise from the probe cables is minimized.
•
The patient must not speak or move the jaws.
•
The patient must breathe calmly.
Physiological conditions
•
Variations that are in synchrony with the patient's pulse:
In some patients, superficial blood vessels may influence the measurement. Try to place the probe at a different
depth in the ear canal. If the patient in general is troubled by a pulsatile tinnitus, this may indicate that further con-
sultation should be done to rule out a vascular middle-ear pathology.
•
Variations that are in synchrony with the patient's breathing:
This is a sign of Eustachian tube dysfunction and may .prevent further reflex testing.
•
Variations due to a very steep tympanogram (for example seen in excessively flaccid tympanic membranes):
This problem can usually be remedied by slightly offsetting the air pressure from the tympanometric peak pressure
(TPP). In case of a negative TPP, it is recommended that the pressure is offset further in the negative direction. In
case of a positive TPP, it is recommended that the pressure is offset further in the positive direction.
An offset of approximately 20 daPa is usually sufficient to stabilize the baseline. Note that excessive offsets will
decrease the threshold sensitivity by several dB.
Use the pressure offset button
±P
to select a pressure offset to stabilize the eardrum. Keep the button depressed
and use the dial to select the desired pressure.
Reflex Threshold results table
Use the buttons in the Reflex section on the front panel to navigate between the measurements. Your selection is high-
lighted in the Reflex results table.
The results table shows the results related to the currently selected curve:
•
Stimulus side
•
Stimulus frequency
•
Threshold levels
62
Otometrics - MADSEN Zodiac
9 Testing with Zodiac Diagnostic or Clinical - Stand-Alone
Содержание zodiac
Страница 1: ...MADSEN Zodiac Reference Manual Doc No 7 50 1610 EN 02 Part No 7 50 16100 EN...
Страница 21: ...4 2 Bottom view A Probe connections B Label Otometrics MADSEN Zodiac 21 4 Views of Zodiac...
Страница 156: ...jB jB App 2 5 2 G App 2 5 3 B G B G Y B G B G 156 Otometrics MADSEN Zodiac App 2 MADSEN Zodiac...
Страница 157: ...App 2 5 4 226 Hz 226 Hz 200 daPa Otometrics MADSEN Zodiac 157 App 2 MADSEN Zodiac...
Страница 206: ...152 151 151 152 150 152 206 Otometrics MADSEN Zodiac Index...