MadgeTech RFC1000 Скачать руководство пользователя страница 14

Product User Guide

RFCurrent2000A: 

To program the channel on the RFCurrent2000A data logger, start by switching 

the wireless mode to 

OFF

 by holding down the 

Wireless

 button on the data logger for 5 seconds.  

Step 1:

 Use the USB Cable, plug the USB end of the cable into an available USB port on the PC.

Step 2:

 Plug the opposite end of the cable into the communication port on the RFCurrent2000A.

Step 3:

 Open the MadgeTech Software. Locate and select the RFCurrent2000A in the 

Connected 

Devices

 panel. 

Step 4:

 In the 

Device

 tab, click the 

Properties

 icon. The 

Properties

 screen will display information 

about the device including 

Wireless

 setting. 

Step 5:

 Under the 

Wireless

 tab, select a desired channel (11-25) that will match with the RFC1000. 

Save

 all changes, disconnect the data logger, and return the device to wireless mode by 

holding down the 

Wireless

 button for 5 seconds. 

Deploying and Activating Devices

Step 1: 

Plug the RFC1000 into the USB port on the base station computer. 

(Additional RFC1000s  

can be used as repeaters to transmit over greater distances)

Step 2: 

If using multiple RFC1000s plug each one into a wall outlet in the desired locations.  

(If transmitting over a distance greater than 1000 feet indoors or 4000 feet outdoors or there 
are walls/obstacles/corners that need to be maneuvered around, set up additional RFC1000s  
as needed.)

Step 3: 

Verify that the data loggers are in wireless transmission mode. 

(See Channel Programming 

steps above)

Step 4: 

On a Windows PC, launch the MadgeTech 4 software program. All active data loggers will 

be listed in the software showing that the device(s) are recognized. 

Step 5: 

To activate the data loggers, click on one to highlight, then click the 

Claim

 icon, and then 

click the 

Start

 button. Do this for each logger in the list that needs to be activated.

RFC1000 General Specifications

12

Interface Type USB (to PC) / Wireless (to Data Logger)

Operating Environment -20 °C to +85 °C, 0 %RH to 95 %RH non-condensing

LED Indicators Red & Green

Enclosure Materials ABS Plastic (body), PVC Plastic (antenna)

Dimensions Enclosure: 3.8 in x 1.6 in x 0.8 in / Antenna: 7.2 in

Approvals FCC ID: OA3MRF24J40MC, IC#: 7693A-24J40MC

Transmission Distance 

(To other RFC1000’s)

•  RFC1000 & RFC1000-IP69K 

4,000 ft max. outdoors - line of sight unobstructed 
1,000 ft max. indoors - typical urban environment

•  RFC1000-CE 

2,500 ft max. outdoors - line of sight unobstructed  
700 ft max. indoors - typical urban environment

Transmission Distance 

(To data loggers)

•  RFC1000, RFC1000-CE & RFC1000-IP69K 

2,000 ft max. outdoors - line of sight unobstructed 
500 ft max. indoors - typical urban environment

Maximum number of  

connected data loggers

64

Frequency 2.405 GHz - 2.475 GHz

Ingress Protection IP40

Содержание RFC1000

Страница 1: ...RFCurrent2000A To view the full MadgeTech product line visit our website at www madgetech com RFCurrent2000A Wireless Current Data Logger Product User Guide ...

Страница 2: ...tional Features Operation 3 Mounting Instructions 4 Product Maintenance 5 Recalibration 5 RMA Instructions 5 General Specifications 6 7 RFC1000 Transceiver Product Overview 9 Installation Guide 9 Mounting Instruction 10 Channel Programming 11 Deployment Activation 12 General Specifications 12 Compliance Information 13 Troubleshooting Tips Back Cover Table of Contents ...

Страница 3: ...e Start icon 7 To download data select the device in the list click the Stop icon and then click the Download icon A graph will automatically display the data Quick Start Steps 1 Install the MadgeTech 4 Software and USB Drivers onto a Windows PC 2 Before plugging the data logger in the computer make sure the wireless setting on the device is set to Wireless OFF If wireless mode is on press and hol...

Страница 4: ...FC1000 wireless transceiver to transmit data back to a central PC The RFCurrent2000A transmits up to 500 ft line of sight to the RFC1000 transceiver Multiple RFC1000 s can be used as repeaters and transmit up to 1000 ft line of sight to other RFC1000 s This allows the data to be sent over a greater distance Selection Buttons The logger is designed with 3 direct selection buttons Scroll Units and W...

Страница 5: ...have met before notifications occur and select which notifications will occur from the drop down menus 4 If selecting to receive notifications by email or text message click Email settings and make sure the correct info has been entered Email and text message notifications will not be received if settings are incorrect 5 Click OK The OK button will be disabled if the Rule name box is empty Changin...

Страница 6: ...ottom of the logger snaps in place to use and mount the unit on a table top or horizontal surface The base can also securely snap to the backside of the logger for wall mounting if desired as shown below There are two holes in the base to allow for screws for mounting Mounting Instructions 4 ...

Страница 7: ...the blank fields 3 Complete the applicable fields on the form including customer Billing and Shipping information even if they are the same Please see the field explanation below for a more detailed description about questions asked in the Device Information section 4 When all of the fields are complete click Generate RMA 5 Print out the confirmation page that follows containing the RMA number and...

Страница 8: ... Date Automatically recorded through the device Battery Type Ultralife 9V Lithium included user replaceable Battery Life Approximately 4 years typical for 15 minute reading rate Data Format µA mA A Time Accuracy 1 minute month Computer Interface USB to mini USB 250 000 baud for standalone operation or RFC1000 RFC1000 CE RFC1000 IP69K required for wireless operation Software XP SP3 Vista Windows 7 ...

Страница 9: ... Battery Warning WARNING BATTERY MAY LEAK FLAME OR EXPLODE IF DISASSEMBLED SHORTED CHARGED CONNECTED TOGETHER MIXED WITH USED OR OTHER BATTERIES AND OR EXPOSED TO FIRE HIGH TEMPERATURE DISCARD USED BATTERY PROMPTLY KEEP OUT OF REACH OF CHILDREN Specifications subject to change See MadgeTech s terms and conditions at www madgetech com 7 Sensor Type Current Sensing Circuitry Nominal Range 20 mA 160 ...

Страница 10: ...Product User Guide RFC1000 Wireless Transceiver 8 Other models available RFC1000 CE Approved for European markets RFC1000 IP69K Splash Proof for harsh environments ...

Страница 11: ... condensing The RFC1000 is rated IP40 and is protected against solids that are greater than 1mm in size This device is not water resistant LEDs The red LED indicates that the device has power The green LED will blink when communicating with other devices Installation Guide Installing the MadgeTech 4 Software Insert the Data Logger Software Flash Drive into an open USB port on a Windows PC If the a...

Страница 12: ...wireless performance both the RFC1000 and the data loggers should be mounted in the same orientation This usually means that the external antenna should be pointing straight up The antenna can pivot to accommodate either a wall mount or a desk mount 10 ...

Страница 13: ...ata logger or RFC1000 that is on the same network is required to use the same channel If all of the devices are not on the same channel the devices will not communicate with one another The images below show the orientations available of the switches for each channel Channel 26 all switches in the up position is not supported 11 Follow the instructions below to configure the channel settings of th...

Страница 14: ... or there are walls obstacles corners that need to be maneuvered around set up additional RFC1000s as needed Step 3 Verify that the data loggers are in wireless transmission mode See Channel Programming steps above Step 4 On a Windows PC launch the MadgeTech 4 software program All active data loggers will be listed in the software showing that the device s are recognized Step 5 To activate the dat...

Страница 15: ...pplicables aux appareils radio exempts de licence L exploitation est autorisée aux deux conditions suivantes 1 l appareil ne doit pas produire de brouillage et 2 l utilisateur de l appareil doit accepter tout brouillage radioélectrique subi même si le brouillage est susceptible d en compromettre le fonctionnement Under Industry Canada regulations this radio transmitter may only operate using an an...

Страница 16: ...software click the File button then click Options 2 In the Options window click Communications 3 The Detected Interfaces box will list all of the available communication interfaces If the RFC1000 is listed here then the software has correctly recognized and is ready to use it Check that Windows recognizes the connected RFC1000 wireless transceiver If the software does not recognize the RFC1000 the...

Отзывы: