QuadTemp
4-Channel Thermocouple Based Temperature Recorder
Red 485
Dark Blue Pantone 289
Light Blue 58% of Pantone 289
Product Information Card
QuadTemp and OctTemp
OctTemp
8-Channel Thermocouple Based Temperature Recorder
Product Quick Reference Card
Part Number
QuadTemp
OctTemp
Internal Channel
Temperature Sensor
Semiconductor
Internal Channel
Temperature Resolution
0.05°C
Internal Channel Accuracy
+0.5°C
Remote Channel Temperature
Sensor Range, Resolution & Accuracy
*See Table for Details
Cold Jct. Compensation
Automatic
Channels
1 internal & 4 remote
1 internal & 8 remote
Memory
500,000/channel
500,000/channel
Sample Rate
4 seconds up to 24 hours
LED Indicator
None
Required Interface Package
IFC110 or IFC200
Baud Rate
115,200
Typical Battery Life
18 months typical
Operating Environment
–20 to +60°C, 0 to 95%RH (Non–Condensing)
Material
Anodized aluminum
Dimensions
7.24” x 2.7” x 1.02”
(183mm x 68mm x 26mm)
7.24” x 2.7” x 1.26”
(183mm x 68mm x 32mm)
Approvals
CE
Specifications subject to change.
See MadgeTech’s terms and conditions at www.madgetech.com
Device Maintenance
Battery Replacement
Materials:
3/32” HEX Driver (Allen Key)
Replacement Battery (U9VL-J)
- Remove the cover from the device by unscrewing the four
screws.
- Remove the battery from its compartment and unsnap it
from the connector.
- Snap the new battery into the terminals and verify it is secure.
- Replace the cover taking care not to pinch the wires. Screw
the enclosure back together securely.
Note: Be sure not to over tighten the screws or strip the threads.
Battery Warning
WARNING: FIRE, EXPLOSION, AND SEVERE BURN HAZARD. DO NOT
SHORT CIRCUIT, CHARGE, FORCE OVER DISCHARGE, DISASSEMBLE,
CRUSH, PENETRATE OR INCINERATE. BATTERY MAY LEAK OR
EXPLODE IF HEATED ABOVE 60°C (140°F).
Recalibration
The QuadTemp or OctTemp standard calibration is one point
at 25°C for the internal temperature sensor and 0mVs for the
thermocouple channels.
Pricing:
Recalibration traceable to NIST
$60.00
Recalibration
$40.00
Additional:
As Found Data
$15.00 per parameter/channel
Verification Point
$15.00 per point
Channel (1st)
$30.00 at 25°C
$45.00 at custom point
Additional Channels
$3.00 at 25°C
$4.50 at custom point
To send the devices back, visit www.madgetech.com, select
Services
then
RMA Process
.
* Remote Channel Range, Resolution & Accuracy
Thermocouple
Range (°C)
Resolution
Accuracy
J
-210 to +760
0.1°C
+0.5°C
K
-260 to +1370
0.1°C
+0.5°C
T
-260 to +400
0.1°C
+0.5°C
E
-260 to +980
0.1°C
+0.5°C
R
-50 to +1760
0.5°C
+2.0°C
S
-50 to +1760
0.5°C
+2.0°C
B
+60 to +1820
0.5°C
+2.0°C
N
-260 to +1300
0.1°C
+0.5°C