Manual
Instructions for use
Thermoplastic Masks
9
26042021V21IFU117000
MacroMedics ®
5.6
R
emoving the Profiles
I.
Always start with the head Profiles, then the shoulders.
II.
Press on the middle of the Profile.
III.
Use your other hand to release all Plungers.
Do not pull too hard! This will lock the Profile.
IV.
Set the plungers in released position during storage.
When removing the profiles becomes difficult:
•
Clean the baseplate holes with a cotton bud.
•
Spray Teflon spray in the hole.
•
Remove residual teflon spray from the hole.
PUSH
PULL
PULL
Released
Position
Locked
Position
Incorrect
Position (locked)