background image

Manual

Instructions for use

DSPS

-

Prominent

SOLUTIONS IN RADIOTHERAPY

6

19092022V2IFU113000

MacroMedics ® 

5.1 

       

MOULDING THE SUPINE OCCIPITAL MASK 

1. 

Make sure that the rotatble lock pawls on either side

              of the frame are in ‘unlocked’ positions.

2.  

Rotate the top part of the mouldroom tool to one side so that it is not 

 

in the way when the patient has to lie down.  

3. 

Take a new supine occipital mask (dark blue, RayOC thermoplastic mask 

 

with the black frame).

4.a 

When using a Waterbath (of min. internal dimensions, in mm: 665w x 575d x 80h - we recommend 

 

using the MacroMedics MacroBath 4)

 to heat the mask, set the water temperature between 68°C 

 

and 70°C (154.4°F and 158°F). Leave the mask in the water for approximately 15 to 20 seconds until 

 

it becomes sufficiently pliable. Take the mask out of the waterbath and dry it between two towels. 

 

Check the temperature of the mask before skin contact.

4.b 

When using an Oven

 to heat the mask, set the oven temperature between 68°C and 70°C 

 

(154.4°F and 158°F). Leave the mask in the oven for approximately 2 minutes and 50 seconds to 

 

3 minutes, until it becomes sufficiently pliable. Take the mask out of the oven and check the 

 

temperature of the mask before skin contact. If the mask is too warm, let it cool down before 

 

placing it on the patient.

5. 

Ask the patient to sit up. Place the mask onto the DSPS-

Prominent

 cradle so that the pins at the 

 

bottom of the black frame fall into the designated holes of the cradle.

6. 

Bend both the ends of the frame (at shoulder height) along the slope of the cradle, 

 

lock the frame’s ends with the rotatable lock pawl that is positioned closest to the frame by 

 

turning it until it is locked, as shown below. 

7. 

Ask the patient to gently lie down onto the mask. Gently position the patient’s head on top of the 

 

mask in the desired position for treatment.  

 

The mask will sink due to the weight of the head and shoulders. The back of the head should just 

 

touch the bottom of the mouldroom tool, but not rest on it. If the depth of the impression is 

 

not deep enough, gently apply pressure with the palm of your hands on the patient’s forehead 

 

to achieve the prescribed position and depth for the mask. Swing the top part of the mouldroom 

 

tool above the patient’s head to check if the chin fits underneath it. If it does not, reposition the 

 

angle of the head slightly while the occipital mask is still mouldable. 

Lock 

Unlock 

Frame

locked 

Frame

unlocked 

Max. chin height

Содержание DSPS-Prominent

Страница 1: ...ADIOTHERAPY MacroMedics BV Oostbaan 670 2841ML Moordrecht The Netherlands 31 0 182 389 777 info macromedics com www macromedics com MANUAL 113000 Instructions for use English Instructions MacroMedics...

Страница 2: ...Manual Instructions for use DSPS Prominent SOLUTIONS IN RADIOTHERAPY 2 19092022V2IFU113000 MacroMedics This page is intentionally left blank...

Страница 3: ...THERAPY 3 19092022V2IFU113000 MacroMedics Table of Contents 1 Warnings 4 2 MRI Safety Information 4 3 Intended Use Indications for Use 4 4 Baseplate set up 5 5 Mask set up 5 6 Cleaning Storage and Dis...

Страница 4: ...ting times Ensure that the mask is not too hot for skin contact before placing it on the patient Do not remove the mask from the patient until the material cools completely Removing the thermoplastic...

Страница 5: ...m of the baseplate with the two couchstrips and carefully place the baseplate on the couchstrips 5 Mask set up Follow your department s patient positioning protocol For DSPS treatments there should be...

Страница 6: ...e Take the mask out of the oven and check the temperature of the mask before skin contact If the mask is too warm let it cool down before placing it on the patient 5 Ask the patient to sit up Place th...

Страница 7: ...black sliding lock pawls are in unlocked position as shown here 5 a When using a Waterbath of min internal dimensions in mm 665w x 575d x 80h we recommend using the MacroMedics MacroBath 4 to heat th...

Страница 8: ...er 5 1 by locking the second rotatable lock pawl 11 Swing the top part of the mouldroom tool above the patient s head again to check if the chin still fits underneath it 12 Use both hands to mould the...

Страница 9: ...e black sliding lock pawls Place your thumb on top and your index finger at the bottom of the black lock pawl With care gently push a little upwards with your index finger to unlock it and then pull w...

Страница 10: ...s in locked position Dispose of the device in accordance with the local laws and regulations applicable to your country 7 Warranty The MacroMedics products are warranted to be free from defects in man...

Страница 11: ...for important cautionary information such as warnings and precautions that cannot for a variety of reasons be presented on the medical device itself Prescription Only Devices for sale to physicians o...

Страница 12: ...UTIONS IN RADIOTHERAPY MacroMedics BV Oostbaan 670 2841ML Moordrecht The Netherlands 31 0 182 38 97 77 info macromedics com www macromedics com Visit www macromedics com for a digital copy of this man...

Отзывы: