Step 4 – Pull the Index Plunger outward to fully remove the Inner Knee Pad Pivot Assembly.
Step 5 – Use a 6mm Allen Key to remove the bolt securing the Outer Knee Pad Pivot
Assembly. Grab the Spacer while removing the Bolt
Step 6 – Remove the Outer Knee Pad Pivot Assembly. Take note which hole position the
second plunger was aligned with.
Refitting / Replacement
Step 7 - Refitting is a reversal of the removal process noting the following points:
A)
Apply Loctite 243 to the Bolt securing the Outer Knee Pad Pivot Assembly to the Mast.
B)
Ensure the Outer Knee Pad Pivot Assembly is orientated the correct way, with the Index
Plunger facing away from the Stand Aid.