background image

 

995064 – Revision A  

 

Page 10 of 36 

 

8.0 – Troubleshooting 

 

Should a problem arise with the use of the Stand Aid, this section offers guidance to all known faults 
revolving the product. The list of faults below offers the correct actions in sequence to resolve the 
issue.  

For each step per scenario, if the step does not determine the issue, continue to the following step 
in sequence until the fault is found. Once the fault is found, refer to the action guide with the 
corresponding number to resolve the issue. After following the correct action, check that the Stand 
Aid is now functioning correctly, and perform a simple test. See Section 4.0 for correct test 
procedure. 

 

For guidance on how to dismantle the hoist to perform troubleshooting actions, see section 9.0 – 
Servicing. 

If the fault is not found and/or the solutions do not correct the problem, contact your local 
Mackworth authorized dealer immediately – contact details are provided on the last page of this 
manual. 

8.1 – Stand Aid doesn’t turn on 

Operate the handset to determine if the Stand Aid powers up. If it does not – Follow the 
Troubleshooting guide below. 

Troubleshooting Guide 

1.

 

Check the E-Stop button on the control box has not been activated.  

2.

 

Check the Handset is correctly connected to the correct control box port. 

3.

 

Check the Handset lead is not damaged, e.g. cuts and breaks. 

4.

 

Check to see if the battery is out of charge and requires charging. 

5.

 

Check the Battery is located correctly on top of the control box.  

6.

 

Check for any damage to the battery terminals, e.g. burns and breaks. 

7.

 

Perform a visual check on the Control box for any damage. 

8.

 

Replace current Control Box. 

Action Guide 

1.

 

Twist the E – Stop button in a clockwise direction to reset. 

2.

 

Re-attach the Handset to the correct control box port – See Section 9.2.1 on Handset 
attachment. 

3.

 

Remove the Handset and Replace –See Section 9.2.1 on Handset attachment. 

4.

 

Plug the charging lead into a wall socket see if the Stand Aid begins to Charge.  The control 
box should show a steady orange LED to indicate the Stand Aid is now charging. If 
unsuccessful, continue Troubleshooting Guide and refer to the section on “Stand Aid 
doesn’t Charge” as this is likely the guide to resolve the issue.
 If Stand Aid turns on, allow 
the Stand Aid to charge before testing. If the Stand Aid does show signs of charging but does 
not turn on, continue this Troubleshooting Guide. 

5.

 

Re-attach the Battery in the correct fashion shown in section 9.1. 

6.

 

Remove the Battery and Replace as shown in section 9.1 

7.

 

Replace the Control Box – See section 9.3 

8.

 

See section 9.3 on Control Box replacement. 

Содержание SA180E

Страница 1: ...Doc No 995064 Revision of document A Rev Date 26 04 2021 SA180E Stand Aid Service Manual...

Страница 2: ...functioning but LCD and or LED is inactive 12 8 6 Stand Aid E Stop button not working 12 8 7 Stand Aid has Power but does not respond to Handset commands 13 8 8 Troubleshooting Hoist Safety Mechanism...

Страница 3: ...er 2 0 Safety Precautions Read and understand this manual in its entirety before servicing the Mackworth Essentials SA180E Stand Aid Mackworth authorised personnel must complete the Full and Interim s...

Страница 4: ...s to see if they are loose if so tighten accordingly Check all pivot points for wear Ensure all Hand Knobs are functioning as intended and without damaged thread Ensure Handgrips are attached and or i...

Страница 5: ...cks are present and the Foot Tray is attached correctly Ensure Foot Tray is fitted correctly Manual Leg Opener Check Leg Opener device opens and closes the Legs as intended ensure it does not require...

Страница 6: ...er Assembly using lift straps 8 Raise the Weight 2 3 the lifting range Be vigilant for unusual noises while raising the weight 9 Deactivate the brakes and close the legs using the handset 10 Drive the...

Страница 7: ...owing Tools Required 3mm Allen Key 4mm Allen Key x2 5mm Allen Key x2 6mm Allen Key x2 8mm Allen Key 17mm Spanner x2 17mm Ratchet Spanner Flat Head Screwdriver Circlip Pliers Pocket Knife Equipment Req...

Страница 8: ...current laws for disposal within the community the device is being used within If there is any uncertainty of the below guidelines contact your local authorities to determine the proper method of disp...

Страница 9: ...07903 RH Leg Assembly 107260 Foot Tray Assembly 107915 Lift Actuator Kit 107308 Knee Pad Assembly 107901 Front Castor Kit Pair 107902 Rear Castor Kit Pair 107210 Mackworth Essentials SA180E Base Assem...

Страница 10: ...tivated 2 Check the Handset is correctly connected to the correct control box port 3 Check the Handset lead is not damaged e g cuts and breaks 4 Check to see if the battery is out of charge and requir...

Страница 11: ...nt 4 Connect the Battery correctly to the Control Box as shown in section 9 1 5 Replace the Battery see section 9 1 6 Replace the Control Box see section 9 3 7 See section 9 2 3 on Charger Lead replac...

Страница 12: ...ift Actuator replacement 2 See section 9 3 for guidance on Control Box replacement 8 5 Stand Aid is functioning but LCD and or LED is inactive Attempt to raise lower the Stand Aid with the handset the...

Страница 13: ...Re attach the Handset to the Control Box See Section 9 2 1 on Handset attachment 3 Re attach the Lift Actuator Cable to the Control Box See section 9 2 2 4 Remove the Handset from the Hoist and Replac...

Страница 14: ...E NOTE Before carrying out any work on a Mackworth Essentials SA180E Stand Aid remember to remove the charger lead and remove any external power source Prior to any servicing the Emergency Stop Button...

Страница 15: ...ll explain the correct procedure on removing refitting the Cables for servicing procedures or replacement 9 2 1 Cables Handset Removal Step 1 Remove the Handset connecter from the Control Box by pulli...

Страница 16: ...ove the Jack End of the Cable which ports the Control Box by pulling downward This is the middle Control Box Port as shown in the image Step 2 Release the cable from the Actuator by turning the lock m...

Страница 17: ...tting the Cable into the Actuator port ensure the locator aligns with the locator inside the actuator port 9 2 3 Cable Charger Removal Step 1 Where Applicable turn the charger off at the wall plug and...

Страница 18: ...it will explain the correct procedure on removing and reinstalling the Control Box for servicing procedures or replacement Removal Step 1 Remove Battery from the Stand Aid Refer to section 9 1 Step 2...

Страница 19: ...procedures or replacement Removal Step 1 Remove the Battery refer to section 9 1 Step 2 Remove the Cables Leads from the Control Box refer to section 9 2 Step 3 Remove the Control Box refer to section...

Страница 20: ...See section 9 2 2 Step 2 Using two 4mm Allen Keys remove one of the screws and washer from the pin at the Top Actuator fixing point Step 3 Pull the Pin out from the top fixing to release the top end o...

Страница 21: ...efitting is a reversal of the removal process noting the following point A Ensure the Actuator is fitted in the correct orientation With the LIANK Specification label facing towards the Mast B Ensure...

Страница 22: ...e the Lift Actuator Cable See section 9 2 2 Step 2 Remove the Lift Actuator See section 9 5 Step 3 Using two 6mm Allen Keys remove one of the Bolts securing the Upper Assembly to the Frame Step 4 Usin...

Страница 23: ...g point A Ensure the threaded boss is placed within the Upper Assembly when fitting B Ensure Loctite 243 is applied to both bolts when fixing the Upper Assembly C When fitting the Upper Assembly ensur...

Страница 24: ...our Bolts circled in the image below using a 6mm Allen Key Step 4 The Leg Rods can be removed from the product Refitting Replacement Step 5 Refitting is a reversal of the removal process noting the fo...

Страница 25: ...e Stand Aid on its side for easier service Step 3 Remove the Leg Rod Assembly from the Base Refer to section 9 7 Step 4 Remove the Bolt from the Leg Pivot Plate as shown below using a 6mm Allen Key St...

Страница 26: ...Remove the Bolt securing the Leg Pivot Plate to the Base using an 8mm Allen Key Ensure to grab the spacer black washer and white washer Step 9 Remove the Leg Leaver Plate as shown ensure to grab the...

Страница 27: ...the Front Castor for servicing procedures or replacement Removal Step 1 Remove the Foot Tray from the Base Refer to section 9 12 Step 2 Place the Stand Aid in its side for easier service Gently rest...

Страница 28: ...and Aid in its side for easier service Gently rest the Stand Aid on its side to avoid damage such as paint scratching Step 3 To release the Rear Castor place a 6mm Allen Key into the bolt and remove t...

Страница 29: ...to avoid damage such as paint scratching Step 3 Remove the Leg Rod Assembly Refer to section 9 7 Step 4 Remove the Front Castors Refer to section 9 9 Step 5 Using two 5mm Allen Keys remove one of the...

Страница 30: ...the Foot Tray for servicing procedures or replacement Removal Step 1 Foot Tray is attached to the Base by the Pins circled in the image below Where applicable use circlip pliers to remove the retainin...

Страница 31: ...anners remove a nut from either side of the Knee Pad Attachment Bar Ensure to grab the Nylon Washer circled Step 2 Slide the Threaded Bar through the Inner Knee Pad Pivot Assembly to remove the Knee P...

Страница 32: ...ad Pivot Assembly Take note which hole position the second plunger was aligned with Refitting Replacement Step 7 Refitting is a reversal of the removal process noting the following points A Apply Loct...

Страница 33: ...6 D When reattaching the Knee Pad to the Threaded Bar ensure the Washers are fitted in the correct positions Nut washer nylon washer Knee Pad nylon washer Inner Knee Pad Pivot Assembly nylon washer Kn...

Страница 34: ...lly off the Base to remove Step 4 Remove the Foot Tray from the Base See section 9 12 Step 5 Place the Base on its side for easier service Step 6 Remove the Rear Castors See section 9 10 Step 7 Remove...

Страница 35: ...ol Box See section 9 2 Step 3 Remove the Control Box from the Mast See section 9 3 Step 4 Remove the Attachment Bracket from the Mast See section 9 4 Step 5 Remove the Lift Actuator from the Stand Aid...

Страница 36: ...ntained in this assembly and installation manual no liability can be accepted by Mackworth for any errors or omissions Mackworth operates a policy of continuous improvement Specifications and other da...

Отзывы: