999061 – Revision A
P a g e
|
18 of 28
9.0
FAULT FINDING
Should a problem arise with the use of the Transfer Aid, review the Table below.
Find the fault and complete the recommended solution.
If the fault is not found and/or the solution does not correct the problem, contact your local Prism Medical UK authorized dealer
immediately – contact details are provided on the last page of this manual.
Fault
Action
Castors do not operate.
Ensure the brake has been fully disengaged, Check the castors are free running
Check the castors are touching the ground. If this does not correct the fault, then
contact your local authorised dealer immediately so the Transfer Aid can be
checked to ensure proper continued operation.
Knee Pad does not lock into position
Ensure that the index plunger is aligned with the holes in the handle.
If this does
not correct the fault, then contact your local authorised dealer immediately so
the Transfer Aid can be checked to ensure proper continued operation.
Handle is loose
Ensure that the star nuts are fully tightened (Refer to section 4.0)
If this does not
correct the fault, then contact your local authorised dealer immediately so the
Transfer Aid can be checked to ensure proper continued operation.
Brakes do not operate
Ensure that the brake mechanism is fully engaged, check to see if both wheels
are touching the ground. If this does not correct the fault, then contact your
local authorised dealer immediately so the Transfer Aid can be checked to
ensure proper continued operation.
Table 8
10.0
GENERAL INSPECTION, MAINTANANCE AND
CLEANING
10.1
Service
No service is to be carried out on the Raiser Transfer Aid while transferring a person to reduce the risk of injury.
Service must be completed by a Prism Medical UK authorised Service Engineer.
Do not attempt to service the product yourself, or warranty is void.
To ensure the safety and continued good function of your Transfer Aid, routine service must be performed on your Raiser
Transfer Aid.
Service should be completed by a Prism Medical UK approved service engineer every 6 months to ensure the products required
standard is maintained. The service history of the product should be documented each service in the Service Log at the back of
this User Manual.
When the Transfer Aid is serviced, the 6 month service checklist must be completed for the Raiser Transfer Aid.
Service Manual Document Number: 995061.
Spare Parts Manual Document Number 992061
The Service must be completed every 6 months after installation of the Transfer Aid to comply with LOLER Regulations.
The Raiser Transfer Aid has an expected Service Life of 10 Years.
Contact your local authorised Prism Medical UK dealer if you:
•
Need more information.
•
Have any questions about the use or service of your Transfer Aid.
•
Notice any change in the performance.
•
Want to report an unexpected occurrence.
•
Want to arrange a service.
•
Need to ascertain necessary information for replacement parts and components.
Contact details of your local Harvest Healthcare dealer are shown on the last page of this manual.
Содержание Raiser
Страница 1: ...Doc No 999061 Revision of document A Revision Date 19 04 21 Raiser Transfer Aid User Manual...
Страница 15: ...999061 Revision A P a g e 15 of 28 6 2 Raiser Dimensions Double Knee Pad Figure 31 Figure 32 Figure 33...
Страница 23: ...999061 Revision A P a g e 23 of 28...
Страница 24: ...999061 Revision A P a g e 24 of 28...
Страница 25: ...999061 Revision A P a g e 25 of 28...
Страница 26: ...999061 Revision A P a g e 26 of 28 User notes...
Страница 27: ...999061 Revision A P a g e 27 of 28 User notes...