8
Test strip reader QUANTOFIX
®
Relax 1.0, 02.2022
Cleaning
NOTICE
For all cleaning work, the unit must be switched off and disconnected from the
power supply.
WARNING
Fire hazard. Do not use flammable detergents or organic solvents to clean the
unit.
NOTICE
Do not clean the device with acetone or similar products.
1. Wipe the outside of the device with a damp
cloth, if necessary with a mild cleaner, and
then dry the device with a soft cloth. Make
sure that no moisture enters the device or
the printer. Clean the touchscreen with a
lint‑free cloth.
2. Switch off the device and pull out the strip
holder to remove it from the device.
3. Rinse the strip holder with clear water.
Alcohol‑based disinfectants can be used
for disinfection.
4. Press the strip holder completely onto the
metal strip with appropriate force.
5. Push the strip holder back into the unit.
For details on cleaning and maintenance, refer to the complete manual.
Quality control
Each facility must implement its own QC policy.
To check the performance of the test strips and the instrument, we recommend measuring
several standard solutions with known concentration within the measuring range.
For additional assurance of your results, you can also use special control strips for the
instrument's test mode (see operating instructions for more information).