WATER TANK
Before servicing the water tank with the MAC155 Probe in place,
turn off the power and remove the power cable from the unit to
eliminate any possible contact with hazardous voltage.
The MAC155 Probe calibration system uses water that is stored in the water
tank to generate reference gases of a known percent of water vapor by volume.
The water tank holds enough water to last for over a year under normal
operating conditions. Do not wait for the low water level detection circuit to trip.
Instead, a regular maintenance schedule should be set up. The tank should be
drained, flushed, and refilled at least once per year.
Read the “OPERATION” section in this manual for information about filling
the tank.
When it is time to re-fill the water tank, proceed as follows:
1. Remove all electrical power from the instrument, and then open the
cover.
2. Loosen the mounting clamp on the barrel of the MAC155 Probe just
enough to rotate the instrument clockwise approximately 15 degrees.
This will assure that water will not run into the instrument when filling
the tank.
3. Siphon water from the tank as previous described.
4. Flush the tank as previously described.
5. Turn the instrument until it is approximately level and re-fill the tank
with distilled water as previously described. Install the fill plug and
make sure it is tight. Wipe up any water that found its way into the
instrument during this process.
Содержание MAC155H
Страница 2: ......
Страница 4: ......
Страница 9: ... Copyright MAC Instruments 2005 All Rights Reserved MAC155 Rev 12 Pg v 127 ...
Страница 13: ... Copyright MAC Instruments 2005 All Rights Reserved MAC155 Rev 12 Pg ix ...
Страница 15: ... Copyright MAC Instruments 2005 All Rights Reserved MAC155 Rev 12 Pg xi ...
Страница 16: ......
Страница 24: ...MAC155 Rev 12 Copyright MAC Instruments 2005 All Rights Reserved Pg 8 ...
Страница 38: ......
Страница 48: ...MAC155 Rev 12 Copyright MAC Instruments 2005 All Rights Reserved Pg 32 Notes ...
Страница 58: ...MAC155 Rev 12 Copyright MAC Instruments 2005 All Rights Reserved Pg 42 ...
Страница 70: ...MAC155 Rev 12 Copyright MAC Instruments 2005 All Rights Reserved Pg 54 ...
Страница 101: ... Copyright MAC Instruments 2005 All Rights Reserved MAC155 Rev 12 Pg 85 Figure 66 MAC155 Time Setup Menu ...
Страница 104: ...MAC155 Rev 12 Copyright MAC Instruments 2005 All Rights Reserved Pg 88 Drawings Mounting Outline ...
Страница 105: ... Copyright MAC Instruments 2005 All Rights Reserved MAC155 Rev 12 Pg 89 ...
Страница 106: ......
Страница 107: ... Copyright MAC Instruments 2005 All Rights Reserved MAC155 Rev 12 Pg 91 Typical Installation ...
Страница 108: ...MAC155 Rev 12 Copyright MAC Instruments 2005 All Rights Reserved Pg 92 ...
Страница 109: ... Copyright MAC Instruments 2005 All Rights Reserved MAC155 Rev 12 Pg 93 Mounting Adapters ...
Страница 110: ...MAC155 Rev 12 Copyright MAC Instruments 2005 All Rights Reserved Pg 94 ...
Страница 111: ... Copyright MAC Instruments 2005 All Rights Reserved MAC155 Rev 12 Pg 95 Wall Mount Remote Sampling Probe Outline ...
Страница 112: ...MAC155 Rev 12 Copyright MAC Instruments 2005 All Rights Reserved Pg 96 ...
Страница 114: ...MAC155 Rev 12 Copyright MAC Instruments 2005 All Rights Reserved Pg 98 ...
Страница 116: ...MAC155 Rev 12 Copyright MAC Instruments 2005 All Rights Reserved Pg 100 ...
Страница 117: ... Copyright MAC Instruments 2005 All Rights Reserved MAC155 Rev 12 Pg 101 Controller Outline ...
Страница 118: ...MAC155 Rev 12 Copyright MAC Instruments 2005 All Rights Reserved Pg 102 ...
Страница 120: ...MAC155 Rev 12 Copyright MAC Instruments 2005 All Rights Reserved Pg 104 ...
Страница 121: ... Copyright MAC Instruments 2005 All Rights Reserved MAC155 Rev 12 Pg 105 Probe Main Board Layout ...
Страница 122: ...MAC155 Rev 12 Copyright MAC Instruments 2005 All Rights Reserved Pg 106 ...
Страница 123: ... Copyright MAC Instruments 2005 All Rights Reserved MAC155 Rev 12 Pg 107 Controller Main Board Layout ...
Страница 124: ...MAC155 Rev 12 Copyright MAC Instruments 2005 All Rights Reserved Pg 108 ...
Страница 125: ... Copyright MAC Instruments 2005 All Rights Reserved MAC155 Rev 12 Pg 109 PLC Interface Wiring Diagram ...
Страница 126: ...MAC155 Rev 12 Copyright MAC Instruments 2005 All Rights Reserved Pg 110 ...
Страница 128: ...MAC155 Rev 12 Copyright MAC Instruments 2005 All Rights Reserved Pg 112 ...
Страница 129: ... Copyright MAC Instruments 2005 All Rights Reserved MAC155 Probe MAC155 Rev 8 Pg 113 Probe to Controller Wiring ...
Страница 130: ...MAC155 Rev 12 Copyright MAC Instruments 2005 All Rights Reserved Pg 114 ...
Страница 131: ... Copyright MAC Instruments 2005 All Rights Reserved MAC155 Probe MAC155 Rev 8 Pg 115 Probe to Controller Wiring 230V ...
Страница 132: ...MAC155 Rev 12 Copyright MAC Instruments 2005 All Rights Reserved Pg 116 ...
Страница 134: ...MAC155 Rev 12 Copyright MAC Instruments 2005 All Rights Reserved Pg 118 ...
Страница 136: ...MAC155 Rev 12 Copyright MAC Instruments 2005 All Rights Reserved Pg 120 ...
Страница 138: ...MAC155 Rev 12 Copyright MAC Instruments 2005 All Rights Reserved Pg 122 ...
Страница 140: ...MAC155 Rev 12 Copyright MAC Instruments 2005 All Rights Reserved Pg 124 ...
Страница 141: ... Copyright MAC Instruments 2005 All Rights Reserved MAC155 Probe MAC155 Rev 8 Pg 125 Probe Junction Box Outline Drawing ...
Страница 142: ...MAC155 Rev 12 Copyright MAC Instruments 2005 All Rights Reserved Pg 126 ...