Calibration

Button

LCD Display

ON/OFF

Button

Fig. 7: Location of Calibration Button

Fig. 8: Re-insert Switch into its original position

Instruction Manual

MAN-040

11

www.macmedical.com

Adjust Switch

1.  Enter Settings Mode by pressing the Calibration Button (located 

on the left, behind the faceplate - Fig. 7).

A. 

If no input occurs within four seconds, the display flashes 

then moves to the next setting.

B.  Press the Calibration Button to view the settings without 

making changes.

C.  Press the ON/OFF button within four seconds to change 

the setting.

DELAY: 5 min. to 12 hr.

DELAY sets the time-out period. The selections begin at 

the currently set time (3 hours, default).
When the time is less than an hour, the time increases 

by 5 minutes each time you press the ON/OFF button. 

When the time is more than an hour, it increases by 15 

minutes. When the selection passes 12 hours, it restarts the 

selection scrolling at 5 minutes.

Scroll: UP or down (dn)

To have the user’s time override options decrease from a 

longer time to a shorter time, set Scroll for dn (down). To 

increase from a shorter time to a longer time, set Scroll for 

UP.

FLASH: ON or OFF and Sound: ON or OFF

When the setting selection is complete the unit will either:

•  Beep once and display END if the calibration 

button is being held; release the button to 

continue.

•  Beep three times while it displays END. This gives 

you a chance to press the ON/OFF button to 

restart the setting procedure.

If you don’t press either button within three seconds, the 

unit exits the Settings.

2. 

When finished adjusting the switch, place it back into the 

recessed area (Fig. 8). Make sure its wires were not accidently 

disconnected. Align its bolt holes with the holes on the cabinet 

front. Insert the bolts through the cabinet front to secure the 

switch.

3.  Align the holes of the access panel with the interior side panel 

holes and reinsert the bolt to secure the plate to the cabinet. 

Содержание MSCA23618-301

Страница 1: ...Medical Inc Read and understand all of the instructions and safety information in this manual before operating this product INSTRUCTION MANUAL Vented Endoscope Cabinets Interior of Endoscope Cabinet...

Страница 2: ...ustomer Service for assistance This manual also contains general specifications warnings and cautions Table of Contents Warnings and Cautions 3 General Specifications 5 Scope Cabinet Features 6 Unpack...

Страница 3: ...to work on the equipment or the installation of unauthorized parts could result in serious personal injury or result in costly damage Always unplug power cord from power source before attempting any r...

Страница 4: ...Instruction Manual MAN 040 4 www macmedical com This Page Left Blank Intentionally...

Страница 5: ...rying time may vary Custom cabinet physical sizes are also available Call customer service if interested in a custom size Model Vent Style Top Scope Qty Dimensions MSCA23618 301 Ambient Vent Flat Top...

Страница 6: ...cope Hangers 9 22 2 Lower Scope Rack 1 3 Access Panel 1 4 Drain Pan 1 5 Padded Back 1 6 Digital Time Switch 1 7 HEPA Filter 1 8 Fan 1 9 Electrical Box 1 Instruction Manual MAN 040 6 www macmedical com...

Страница 7: ...ng the Endoscope Cabinet Stainless steel Endoscope Cabinets are shipped on well constructed wooden skids Steel bands secure the cabinet to the skid The stainless steel surfaces of the part are protect...

Страница 8: ...oper door functionality Only transport the Cabinet when empty Properly secure while being transported Install the Cabinet in its final position The Cabinet is difficult to move after assembly is compl...

Страница 9: ...e time switcher has four adjustable features time out period time scroll direction time out visual and audible warnings See the Timer countdown chart Fig 4 Time Out Period The timer can be adjusted to...

Страница 10: ...nterior righthand side of the cabinet The switch must be partially removed from the cabinet in order to access the switch Calibration button 1 At the front of the cabinet remove the two bolts that hol...

Страница 11: ...es 12 hours it restarts the selection scrolling at 5 minutes Scroll UP or down dn To have the user s time override options decrease from a longer time to a shorter time set Scroll for dn down To incre...

Страница 12: ...he Filter Access Panel Fig 9 shown in yellow 3 Remove the Filter Access Panel assembly from the air vent panel Fig 10 4 Remove the HEPA filter from the recessed filter box Fig 10 5 Insert a new HEPA f...

Страница 13: ...thoroughly disinfected endoscopes in the provided holders Make sure the endoscopes hang freely so that they are not damaged or contaminated by contact with one another A Remove scope valves and caps a...

Страница 14: ...eneral housekeeping of medical storage cabinets and other non critical items B Quaternary ammonium compounds are bactericidal fungicidal and virucidal against lipophilic enveloped viruses C At hospita...

Страница 15: ...the manufacturer s instruction for proper use of these products Cleaning the Endoscope Cabinet Interior Unplug power vented models to remove the Endoscope Cabinet from its power supply Clean and disin...

Страница 16: ...0461 1836 Drain Pan for 18 x 36 cabinets IC0461 1842 Drain Pan for 18 x 42 cabinets IC0461 2436 Drain Pan for 24 x 36 cabinets IC0461 2442 Drain Pan for 24 x 42 cabinets IC40L 36 Lower Scope Rack 9 IC...

Страница 17: ...N Cleaning Procedures Y N Disinfecting Procedures Y N Date of Cleaning Disinfection Date of Cabinet Inspection Date of HEPA Filter Replacement For your convenience here is a simple chart to use to no...

Страница 18: ...Instruction Manual MAN 040 18 www macmedical com Notes...

Страница 19: ...tion alteration tampering negligence lack of routine maintenance or misapplication or products that have been repaired by other than MAC Medical or its authorized representatives This warranty does no...

Страница 20: ...ll Free 877 828 9975 www macmedical com MAC Medical Inc Printed in USA Publication No MAN 040 Rev H March 2024 Information regarding this product is subject to change without prior notice Customer nee...

Отзывы: