3. 2 Focus adjustment   

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

3.2.1

 Adjusting the rotation tension of the focus 

adjustment knob 

(fig.13

fig.14)

 

        For these Microscopes ,the tension has been 

adjusted well before delivery. 

1

Rotate the focus adjustment knob

  to move 

the slider

  to a lower place for fear that the 

zoom  body  drops  down  suddenly  if  the 

tension is decreased too much.   

2

  Hold

 

the focus adjustment knobs

  with both 

hands, stop the left knob and rotate the right 

knob  to  increase  or  decrease  the  focus  knob 

tension.   

3.  Brightness  adjustment:  the  right  brightness 

adjustment  knob

  is  used  to  adjust  the 

bottom  illumination,  and  the  left  brightness 

adjustment  knob

  is  used  to  adjust  the  top 

illumination 

 

 

3.2.2 Adjusting tension collar

Fig.15

 

 

1.    If the microscope body falls down by its own 

weight  or  the  focus  obtained  by  fine 

focusing  is  lost  immediately,  the  rotation 

tension  adjustment  may  be too light. In this 

case, rotate the  tension collar  with provided 

spanner to increase the rotation tension

..

 

2. The brightness adjustment method is the same 

as that    of    fig14

 

 

 

 

 

 

 

 

Spanner 

Fig.15 

 

 

 

② 

Fig.13 

 

 

Fig.14 

,

Содержание PZ-4

Страница 1: ...l is written for stereomicroscope To ensure the safety obtain optimum performance and to familiarize yourself fully with the use of this microscope it is recommended strongly that you study this manua...

Страница 2: ...2 1 Components 4 2 Assembly 5 2 1 Assembly diagram 5 2 2 Detailed Assembly Procedure 6 3 Adjustment and Operation 9 3 1 Adjusting binocular head 9 3 2 Focus adjustment 10 4 Specification 11 5 Troubles...

Страница 3: ...ottom side as this may result in electric shock 8 Use the factory supplied power cord please 2 Maintenance and Storage 1 All the lenses have been well checked and adjusted It is forbidden to disassemb...

Страница 4: ...wer at once and wipe up the moisture 7 Do not disassemble any parts of the microscope which will affect the function or decline the performance of the microscope 8 Place the instrument in a cool dry p...

Страница 5: ...tereo Microscope Adjustable WF10 eyepieces Focusing Assembly Binocular viewing head Zoom body Clamping Knob Objectives Glass stage Standard base Brightness adjusting knob Power cord Main Switch Coarse...

Страница 6: ...bers indicate the order of assembly When assembling the microscope make sure that all parts are free of dust and dirt and avoid scratching any parts or touching glass surfaces Power cord Microscope ma...

Страница 7: ...ewing head into the mounting port on the zoom body and tighten the viewing head clamping screw with provided hexagonal driver Note Working Environment Requirement 1 Ambient temperature 5 C to 40 C 41...

Страница 8: ...e eyepiece as in Fig 5 When Not Wearing Eyeglasses Extend the folded eyeshades This makes observation easier by preventing the inverse incidence of light from between the eyepiece and your eyes 2 2 4C...

Страница 9: ...depth The size of aperture iris diaphragm can be adjusted Switch the lever to left to open the aperture iris diaphragm to the right to close the aperture iris diaphragm Check the effect during observa...

Страница 10: ...or 6 4 or 10 and turn the focusing knob to focus the specimen 4 Rotate the zoom body knob to the lowest magnification 0 8 looking only into the left eyepiece adjust the diopter ring on left eyepiece...

Страница 11: ...knob to increase or decrease the focus knob tension 3 Brightness adjustment the right brightness adjustment knob is used to adjust the bottom illumination and the left brightness adjustment knob is us...

Страница 12: ...Ratio 1 6 1 8 1 10 Objective Magnification 0 8 5 0 8 6 4 0 8 8 Eyepiece 24 10 WF Diopter adjustable Binocular Viewing Head Inclined at 20 interpupillary distance range 50 75mm Working Distance 78mm V...

Страница 13: ...adjustment ring is too tight Loosen it properly 8 6 Zoom microscope body drops or specimen goes out of focus during observation Tension adjustment ring is too loose Tighten it properly 8 6 Outfit Comp...

Отзывы: