background image

To use Speed controller

1.Turn No.13 dial on the rear side to set the required speed.(No.12)

Note:

1)Speed is controlled from 100 to 1500 rpm by the dial.

  (Turn the dial clockwise, then speedis increased 10 by 10 rpm, and turn the dial counter clockwise, 

then speed is decreased 10 by 10 rpm) 

2)To stop the speed, turn the dial counter clockwise till the end.

3)Set the required speed by dial, then it will arrive at the setting value in a few second and keep the 

   constant speed by arithmetic logic unit.

  (Smooth start function)

4)In initial operating, if the motor does not run, give the motor output fast and make it run until speed 

    is arrive at 10rpm so that the motor runs.

※ When "Err" are shown on display

① Check if the temperature sensor is coneected corretly.

② Check if the wire of the temperature sensor is not down.

③ Check if the tempereture sensor is broken.

④ Check the position of sensor in the sensor switch, internal or external sensor.

       ( It is same situation with No.1 case )

[] If you have any other questions, contact Misung or Misung authorized distributor in your country.

Содержание DMB-1

Страница 1: ...avoid electrical shock always 1 Use a properly grounded electrical outlet of correct voltage and current handling capacity 2 Disconnectfrom power supply before servicing To avoid personal injury 1 Do not use in the presence of flammable or combustible materials fire or explosion may result This item contains components which may ignite such materials 2 Keep the item clean Use non abrasive cleaner ...

Страница 2: ...parameters in parameter set 2 Displays set value and parameter set SV Displays set value in controlling and parameter value in parameter set Press key No 9 then displays the remaining time 3 Lamp of output control OUT 4 Warning lamp of highest and lowest ALM 5 Auto tuning lamp AT Press key No 11 then it starts PID auto tuning with flikering of the lamp and the lamp turns off when it is finish 6 Ti...

Страница 3: ...maticall y set for Auto Tuning 7 I Intergral time Possible to set a 1sec unit within 5 9998 sec range 8 5 ALL Low alarm It sets the lowest alarm limit 0 It is absolute value 4 ALH High alarm It sets the heighest Alarm limit 0 It is absolute value Up key can modulate ON OFF The Timer begins working after current temp reaches set temp After a set time reaches it stops ending Control output The TIMER...

Страница 4: ...e 1 Use without auto tuning but do it in case of a big temperature difference 2 Press No 11 key then it starts with switching on and off 3 Temperature error could be happened on auto tuning but it is controlled correctly when auto tuning is finished 4 Temperature go higher than the setting value first time but it is returned to normal condition immediately Over shoot condition 5 In case that tempe...

Страница 5: ... a few second and keep the constant speed by arithmetic logic unit Smooth start function 4 In initial operating if the motor does not run give the motor output fast and make it run until speed is arrive at 10rpm so that the motor runs When Err are shown on display Check if the temperature sensor is coneected corretly Check if the wire of the temperature sensor is not down Check if the tempereture ...

Отзывы: