STATUS INDICATION

Press check button to indicate the status of internal dis-
charge element.

 

INDICATORS

Activated by CHK (Check) button, see the status table 
shown below. 
BAT: Green LED
ALM: Red LED
CHK button: Push button; momentary

BAT

ALM

BATTERY

DISCHARGE 

ELEMENT

REPLACEMENT

ON

OFF Normal

No Need

ON

ON

Normal

Near End

Near

OFF

ON

Normal

End of Life

Immediately 
Required

OFF

OFF Discharged Unable to Judge

CHECKING

 

WIRING

• Make sure that wiring is done as instructed in the con-

nection diagram. 

• Make sure that the Earth terminal is connected to the 

metallic housing of protected device.

• Make sure that the Earth terminal is connected to earth.

MAINTENANCE

Even lightning in remote locations could induce surges 
without our knowledge.  Regular checking of the surge 
protector is important to find degradations in early stage, 
before and after the storm seasons, and whenever you ex-
perience a strong lightning storm.  Checking procedure is 
explained in the following:

 

CHECKING EXTERIOR

If discoloration or deformation is observed, replace with a 
new one immediately.

 

CHECKING

COMMUNICATION

Check the communication status is improved or not by re-
moving the device when the communication is unstable or 
stopping.  Replace with a new one if the device is not recov-
ered.

DISCHARGE ELEMENT

For testing, press CHK (Check) button on the front and 
LEDs turn on or off according to the surge protector’s condi-
tion.  When replacement is required, replace the module to 
a new one immediately.  Refer to the table for judgement.