M-9500-3.0.6-0 INTEGRA 9500-9505 Operations Manual – English
Approved
: Spike Reed
Effective:
15 Nov 2010
Supersedes:
28 Jan 2010
LVS®
CONFIDENTIAL
All contents copyright © LVS® 2010 Page 34 of 82
The image may appear too light or too dark; this is to be expected. Please do not
attempt to make any adjustments on the lens assembly; it has already been set
at the factory and does not require adjustment. The software automatically
adjusts to the proper brightness level.
STEP 3:
Press the
Calibrate
button. The system begins the calibration
process. The calibration result could be either of the following:
EAN/UPC Calibrated Conformance
Standard Test Card
GS1-128 Calibrated Conformance
Standard Test Card
WARNING:
The Calibration Test Card is used to calibrate the light
source to a known level of white and black. There cannot be any other
item in the field of view when calibration takes place. Failure to comply
with these instructions will distort the final verification grade.
To correctly calibrate, the Calibration Test Card must be flat. Use
magnets or weights, if necessary, to hold the card down flat. The
magnets or weights must not appear in the field of view.