Calibrate the System
STEP 1:
To calibrate the system, click the
Calibration
tab. A Login pop-up
window appears prompting for a user name and password. Enter your user name
and password, and then click the
OK
button.
Note:
If the user has not been granted permission rights to calibrate
the system, an error message appears indicating that the user is not able
to access the Calibration tab.
STEP 2:
Locate the Calibrated Conformance Standard Test Card
assigned to
your system. Choose one of the Master Grade barcodes and place it so the blue
line goes through the middle of the PASS portion of the barcode (see Figure
below). The blue line helps you align the calibration label to the field of view; it
does not mean that only one line is being used to calibrate the system.
Note:
After placing the calibration card in position, check to be sure the
goal values match the values printed on the calibration card. If they do
not match you have the wrong calibration card for the system or you
have to change the goal values. See section “Replacement of
Calibrated Conformance Standard Test Card.”
Place the Master Grade
barcode so the blue line
goes through the middle of
the PASS portion of the
barcode.