7.4.2 Calibration with measuring vessel
One easy way to perform a calibration is by performing a comparison using a sufficiently large and
precise measuring vessel. Proceed as follows:
1. Meter is in dispensation mode, partial volume counter has been
deleted.
2. The dispensation is carried out at a flow speed that is as constant
as possible until the measuring vessel is filled to the defined
volume.
3. Switch to programming mode, sub-menu "Calibration" (skip
volume unit and fluid menus with the "Mode" key).
4. Adjust the calibration factor until the quantity displayed
corresponds to the quantity dispensed into the measuring
vessel.
5. Save the new calibration factor and return to
dispensation
mode
.