Temperature
Manual
0 to 100
℃
, be adjusted by
Compensation
push button on front panel.
for pH
Automatic
With the optional temperature
measurement
( ATC )
probe ( TP-07 )
0 to 65
℃
.
pH
PH7, PH4, and PH10, 3 points calibration
Calibration
ensure the best linearity and accuracy.
Optional
*
PH electrode.......................
PE-03, PE-11, PE-01, PE06HD
probe and
PE-04HD, PE-05T, PE-03K7
accessories
*
ATC ( automatic temperature
probe ).......................................
TP-07
*
pH 7 buffer solution.....................
PH-07
*
pH 4 buffer solution......................
PH-04
*
ORP electrode..............................
ORP-14
Measurement
Range
Resolution
Accuracy
PH
0 to 14 PH
0.01 PH
±(0.02 PH + 2 d)
mV
-1999 to 1999 mV
1 mV
±(0.5% + 2 d)
*
PH accuracy is based on calibrated meter only.
B. Conductivity
Conductivity probe
Optional, Carbon rod electrode for long life.
Function
* Conductivity ( uS, mS )
* TDS ( Total Dissolved Solids, PPM )
* Temperature (
℃
,
℉
)
Temperature
Automatic from 0 to 60
℃
(32 - 140
℉
),
Compensation
with temperature compensation factor
variable between 0 to 5.0% per C.
Probe
0 to 60
℃
.
Operating Temp.
Probe Dimension
Round, 22 mm Dia. x 120 mm length.
Optional
*
Conductivity probe......................................
CDPB-03
probe and
*
Pt Conductivity probe......................................
CDPB-04
accessories
*
1.413 mS Conductivity Standard
Solution...........................................
CD-14
6
Содержание YK-2001PHA
Страница 12: ...3 FRONT PANEL DESCRIPTION Fig 1 10 ...