4-4 Auto power off

In order to prolong the battery life, the instrument
has "Auto Power Off " function. The meter will
switch off automatically if no buttons are pressed
for around 10 minutes.

4-5 Set the Temp. unit (  ,   ) with default

℃ ℉

1)Power OFF the meter.
2)Press the " Hold Button " ( 3-3, Fig. 1 ) continuously,

use another finger power ON  the meter ( press the  "
Power Button " once ), until the Temp. unit (   or    )

is presented on the Display then release the  " Hold
Button ", the Temp. unit will change From   to   ( or

 to   ) with default.

4-6 Zero adjustment

Put the meter into the clean air environment ( not
contain any CO value ), power on the meter by
pressing the " Power Button "  ( 3-2, Fig. 1 ), wait
about 30 seconds to warm up the meter.
If the display not show " Zero " CO value, then pressing
" Alarm Button " ( 3-4, Fig. 1 ) continuously at 
least five seconds will offset the display value and
show " 0  PPM ".

* The zero adjustment procedures are effected only

existing Display value is within 10 PPM.

5

Содержание PCO-350

Страница 1: ...o the field of precision measurement Although this METER is a complex and delicate instrument its durable structure will allow many years of use if proper operating t e c h n i q u e s a r e developed...

Страница 2: ...Sensing Head 3 3 2 Power Button 3 3 3 Hold Button 3 3 4 Alarm Button 3 3 5 LCD Display 3 3 6 Battery Compartment Cover 3 4 MEASURING PROCEDURE 4 4 1 CO measurement 4 4 2 Data hold 4 4 3 CO Alarm sett...

Страница 3: ...ures Auto shut off function to prolong the battery life Operates from DC 1 5V UM4 AAA x 4 PCs batteries Durable and compact ABS plastic housing 2 SPECIFICATIONS 2 1 General Specifications Circuit Cust...

Страница 4: ...ing case CA 52A 1 PC 2 2 Electrical Specifications 23 5 CO Carbon dioxide Range 0 to 1 000 PPM CO Resolution 1 PPM Carbon Accuracy 5 2 PPM monoxide Response 30 seconds time Repeatability 2 Zero drift...

Страница 5: ...3 FRONT PANEL DESCRIPTION Fig 1 3 1 CO Sensing Head 3 2 Power Button 3 3 Hold Button 3 4 Alarm Button 3 5 LCD Display 3 6 Battery Compartment Cover 3...

Страница 6: ...3 3 Fig 1 momentarily to hold the measured value The LCD will show a HOLD symbol 2 Press the Hold Button once again to release the data hold function 4 3 CO Alarm setting Press the Alarm Button 3 4 Fi...

Страница 7: ...the Temp unit or is presented on the Display then release the Hold Button the Temp unit will change From to or to with default 4 6 Zero adjustment Put the meter into the clean air environment not cont...

Страница 8: ...s Environmental Protection Agency 35 PPM Max exposure time 1 hour United States Environmental Protection Agency 50 PPM Max exposure time 8 hour OSHA 100 PPM Max exposure 100 minutes Circulation air UL...

Отзывы: