5)Under the " Hold Button " ( 3-4, Fig. 1 ) and " REC.
Button " ( 3-3, Fig. 1 ) be pressed condition, press
" , Button " ( 3-5, Fig. 1 ) to add 0.1 or press
℃ ℉
" Function Button " ( 3-6, Fig. 1 ) to decrease 0.1 to the
upper digits.
0.0
0.2
6)When desired value available, release the fingers from the
" Hold Button " (3-4, Fig. 1) and "REC. Button " (3-3,
Fig. 1 ) to finish the Offset Temperature Adjustment
Procedures.
CONSIDERATION
The calibration point is highly recommended at
0.0 or normal ambient temperature. Please
never ever adjust the offset temperature over
this range (0.0 to normal ambient temperature).
11