5. OFFSET TEMP. ADJUSTMENT

  Type K THERMOMETER

The instrument is fully calibrated to assure the high

quality measuring performance. Usually it is not

necessary to execute  the Offset Temperature Adjustment
procedures, however the measured result could be
influenced by using different type of  temperature probes.
 Under such condition, customer can easily compensate
the temperature value by push button on the front panel.
However, in general conditions, customers are  always
not recommended to execute the Offset Temperature
Adjustment Procedures without qualified calibration
equipment.

Offset Temp. Adjustment Procedure

1)Power on the instrument.
2)Set the function to the " Type K thermometer "

measurement, refer to page 6, Chapter 4-3.

3)Connect with a reliable temperature calibrator.
4)Adjust the calibrator and make it 0.0 output.
5)Press the " Hold Button " ( 3-4, Fig. 1 ) and " REC.

Button " ( 3-3, Fig. 1 ) at the same time for around 1
seconds, the upper display and lower display will show the
same value like

0.2

0.2

10

Содержание MS-7011

Страница 1: ...x and delicate instrument its durable structure developed Please read the following instructions carefully and always keep this manual within easy reach OPERATION MANUAL Heavy duty probe Dew point Typ...

Страница 2: ...6 3 11 Stand 6 3 12 Probe Plug 6 3 13 Probe handle 6 3 14 Probe main body 6 3 15 Sensing head with inner filter 6 3 16 Protection cover for Sensing head 6 4 MEASURING PROCEDURE 7 4 1 Humidity content...

Страница 3: ...ure linearity and precision cold junction compensation circuit high accuracy Microprocessor circuit assures maximum possible accuracy provides special functions and features Large LCD with two display...

Страница 4: ...ty Content High precision thin film capacitance sensor Temperature Thermistor Type K Thermocouple probe thermometer Data Hold Freeze the display reading by push button Memory Maximum Minimum value Rec...

Страница 5: ...Meter 195 x 68 x 30 mm 7 6 x 2 6 x 1 2 inch Humidity Probe length 600 mm Content Probe head diameter 10 mm Probe Cable length 1 5 meters Accessories Instruction manual 1 PC Included Humidity content p...

Страница 6: ...70 RH 3 RH Range 0 to 50 32 to 122 Temp Resolution 0 1 degree Accuracy 0 8 1 5 Dew Point Range 25 3 to 48 9 Resolution 0 1 Range 13 5 to 120 1 Resolution 0 1 Remark Dew Point display value is calculat...

Страница 7: ...Type K Thermometer Range 100 0 to 1300 0 Resolution 0 1 Accuracy 50 0 to 1300 0 0 2 0 5 50 1 to 100 0 0 2 1 Range 148 0 to 2372 0 Resolution 0 1 Accuracy 58 0 to 2372 0 0 2 1 58 1 to 148 0 0 2 1 8 5...

Страница 8: ...3 5 Button 3 6 Function Button 3 7 Battery Compartment Cover 3 8 Temp Probe input Socket 3 9 Humidity Probe input Socket 3 10 RS 232 Output Terminal 3 11 Stand 3 12 Probe Plug 3 13 Probe handle 3 14 P...

Страница 9: ...3 15 Fig 1 4 Penetrate the Sensing head 3 15 Fig 1 along with Probe main body 3 14 Fig 1 into the measurement materials such as Grain Corn Rice Cotton Paper refer to Fig 2 The meter s display will sho...

Страница 10: ...3 6 Fig 1 the LCD will show the unit DEW or 4 3 Type K thermometer measurement 1 Take out the humidity probe away from the Humidity Probe input Socket 3 9 Fig 1 If not take out the humidity probe the...

Страница 11: ...he maximum value just press the Hold Button 3 4 Fig 1 for a while then the display will show the REC symbol only execute the memory function continuously b Press the REC Button 3 3 Fig 1 again the REC...

Страница 12: ...on the front panel However in general conditions customers are always not recommended to execute the Offset Temperature Adjustment Procedures without qualified calibration equipment Offset Temp Adjust...

Страница 13: ...er digits 0 0 0 2 6 When desired value available release the fingers from the Hold Button 3 4 Fig 1 and REC Button 3 3 Fig 1 to finish the Offset Temperature Adjustment Procedures CONSIDERATION The ca...

Страница 14: ...ata Record function with a REC symbol on the display for long period measurement 7 RS232 PC SERIAL INTERFACE The instrument has RS232 PC serial interface via a 3 5 mm terminal 3 10 Fig 1 The data outp...

Страница 15: ...D9 Decimal Point DP position from right to the left 0 No DP 1 1 DP 2 2 DP 3 3 DP D8 to D1 Display reading D1 LSD D8 MSD For example If the display reading is 1234 then D8 to D1 is 00001234 D0 End Word...

Страница 16: ...ple 40 to 482 Probe Ultra fast response naked bead Model TP 01 thermocouple general purpose application Type K Measure Range 50 to 900 Thermocouple 50 to 1650 Probe Dimension 10 cm tube 8 mm Dia Model...

Отзывы: