The data stream will be displayed in the following
format :
gram
11 digit data stream :
D10 D9 D8 D7 D6 D5 D4 D3 D2 D1
Each digit indicate the following status :
D0
CR
CR = Enter (code is "A")
D1
LF
LF = Line feed (code is "D")
D2
g
D3 to D5
SPACE
D6 to D9
Display reading, D6=LSD, D9=MSD
D10
SPACE
oz
12 digit data stream :
D11 D10 D9 D8 D7 D6 D5 D4 D3 D2 D1
Each digit indicate the following status :
D0
CR
CR = Enter (code is "A")
D1
LF
LF = Line feed (code is "D")
D2
Z
D3
O
D4 to D5
SPACE
D6 to D10
Display reading, D6=LSD, D10=MSD
D11
SPACE
7