Operating Temp. 0 to 50 (32 to 122 ).
℃
℃
℉
℉
Operating
Less than 80% RH.
Humidity
Weight
380 g/0.85 LB (including battery).
Dimension
HWD : 255 x 73 x 38 mm.
(10 x 2.9 x 1.5 inch)
Max. Jaw
51 mm ( 2.1 inch ) Dia.
Open Size
Accessories Operation manual...............................1 PC
Included
Test lead (red & black)........................1 Set
Fuse ( 500 mA, 5 mm dia. x 20 mm )...1 PC
Optional * RS232
cable........................... UPCB-02
Accessories &
* USB cable.............................. USB-01
Adapters
* Data Acquisition software
* Refer to
..................................SW-U801-WIN
page 15
* Carrying case, EMF Adapter,
Light Adapter, Anemometer Adapter,
Pressure Adapter, Sound Adapter,
Tachometer Adapter, High Voltage Probe.
2-2 Electrical Specifications (23 ± 5 )
℃
Function
Range
Reso- Accuracy
Overload
lution
Protection
DC/AC 400
mV
0.1 mV
± ( 0.5 % + 2d )
Voltage
(DC only)
4 V
0.001V
DCV:
40 V
0.01V
± ( 1 % + 2d )
400 V
0.1 V
ACV:
AC/DC 1000 V.
1000V
1 V
± ( 1.2 % + 5d )
DC / AC
400 uA
0.1 uA
Current
4000 uA
1 uA
(Direct input)
40 mA
0.01 mA
± ( 1.2% + 5d )
AC/DC 500 mA
400 mA
0.1 mA
( Fuse )
DC /AC
400 A
0.1 A
± ( 2 % + 5 d )
current
(Clamp on)
2000 A
1 A
± ( 2 % + 8 d )
AC/DC
2000A/1000V
Remark
* True rms measurement both for ACV, ACA function.
* Input impedance for ACV & DCV range is 10 Mega ohm.
* ACA, ACV frequency response is from 45 to 1 KHz.
* ACA, ACV specification be tested on sine wave 50/60 Hz.
2